XNX Universal Transmitter
Warnings/Faults
130
Fault
Description
Appli-
cable
Sensors
Latching / Non-
Latching
Frequency of
Diagnostic
Event History Data
Action For Resolution
F149
Internal 4-20 mA
monitoring circuit
communication
failure
All
Non-latching
3.366 seconds
0
Contact Honeywell Analytics’ Service Department.
F150
mA Output Monitor
Communications
Watchdog Error
All
Non-latching
138 us
Communication error
count
Contact Honeywell Analytics’ Service Department.
F151
Sensor Module
Type Changed
ECC
Non-latching
2 seconds
Module type: 0=None,
1=ECC, 2=mV,
3=Excel, 4=Optima,
5=Generic mA
For ECC: Perform Accept New Sensor function, if problem persists
contact Honeywell Analytics’ Service Department. For others, contact
Honeywell Analytics’ Service Department.
F152
Option Module
Configuration Error
All
Latching
Only at powerup
or every 125 ms
when no option
board detected
Option module ID:
0=None, 1=Founda-
tion Fieldbus, 2=Mod-
bus, 3=Relay
Confirm option properly installed, reconfigure unit.
F153
Signal/Data mis-
match error on IR
personality
Optima,
Excel
Non-latching
2 seconds
Digital sensor reading
Check wiring to Optima/Excel. In particular, check the white wire be-
tween XnX and Optima/Excel. Note: power must be cycled to reset F153
after correcting the cause.
F154
mA Input Diagnos-
tic Failure
Optima,
Excel
Latching
5 minutes after
power up and
then every 8
hours
Input mA
Contact Honeywell Analytics’ Service Department.
F155
Generic mA Sensor
Type Error
Generic
mA
Non-latching
2 seconds
Input mA
Indicates that mA input from sensor is less than 3 mA. Check wiring
between XNX and sensor. Also check the switches S3 and S4 are set
correctly. If the switch settings need to be change, power down the
transmitter before changing the switch settings. If wiring and switches
are okay, replace sensor.
F156
mV Current Control
Failure
mV
Non-latching
Main loop x16
constant current A/D
input mV
Check that correct mV sensor type is selected. Check wiring between
XNX and sensor. If sensor type and wiring are okay, replace sensor.
F157
Sensor Drift Fault
ECC, mV
Non-latching
2 seconds
Current baseline
Perform zero calibration. If problem persists, replace sensor.
F158
Sensor/Personal-
ity Part Number
mismatch
All
Non-latching
"ECC & mV: main
loop x2;
Optima & Excel: 2
seconds"
Entire personality
part #
Check that installed option matches the option indicated in the transmit-
ter part number, check wiring to Optima/Excel.
Summary of Contents for XNX-AM Series
Page 2: ...This page is intentionally left blank ...
Page 7: ...XNX Universal Transmitter Safety and Information 5 Safety and Information ...
Page 12: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 10 Introduction ...
Page 27: ...XNX Universal Transmitter Introduction 25 Menu Navigation ...
Page 102: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 100 Calibration ...
Page 118: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 116 Maintenance ...
Page 122: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 120 Warnings and Faults ...
Page 139: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 137 Specifications ...
Page 174: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 172 Control Drawings ...
Page 177: ...XNX Universal Transmitter Control Drawings 175 ...
Page 179: ...XNX Universal Transmitter Control Drawings 177 Remote Sensor Mount ...
Page 180: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 178 HART Protocol ...
Page 193: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 191 Modbus Protocol ...
Page 200: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 198 Warranty ...
Page 202: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 200 Index ...