XNX Universal Transmitter
Section 3 - Calibration
102
Zero and Span Calibration for EC/mV Sensors and
Searchpoint Optima
Caution:
Before initial calibration, allow the sensor to stabilize for 30 minutes after applying power.
When in zero and span calibration modes, the current output from the sensor is inhibited (default
2mA) to avoid false alarms.
CAUTION
!
For sticky
gases (HCI, HF, CI
2
, CIO
2
, HCN, F
2
and O
3
), use PTFE tubing
with short pieces of
Tygon tube for the final connection (due to the
inflexibility of PTFE). This minimizes adhesion of the gas to the tube
surface and allows more accurate measurement. Use a one-inch section
of Tygon tubing as a union sleeve to join the calibration cup’s fitting and
the PTFE tubing. Push the PTFE tubing against the fitting so they make
secure contact as shown in the illustration. Gas should not be able to
contact the Tygon sleeve. Attach the PTFE tubing to the regulator in the
same manner.
PTFE (Teflon) tubing
calibration cup
calibration gas
stainless steel regulator
EC sensor
Tygon sleeves
Figure 192. Joining the PTFE tubing with Tygon tubing
To calibrate the sensor, use an appropriate span gas cylinder, tubing,
magnet, and calibration gas flow housing. Set the flow regulator to 300-
375 ml/min for XNX EC sensors or 300-700 ml/min for XNX mV sensors.
Use a compressed gas cylinder (20.9%Vol oxygen) to perform the zero
calibration if the sensor is located in an area containing any residual
amount of the target gas. If no residual gas is present, background
air can be used to perform the zero calibration. Contact a Honeywell
Analytics representative for details about suitable calibration kits. To
calibrate the sensor, follow the steps in
Note:
The oxygen sensor does not require a zeroing procedure. Background air (20.9%Vol oxygen)
can be used to span the oxygen sensor in place of a compressed air cylinder (20.9%Vol oxygen). See
the
Sensor Data
section for other sensors.
Note:
EN performance standards require 10 minutes stabilization time for application of zero and span
gas for performance-approved EC, mV, and IR sensors prior to calibration.
NOTE
!
Calibration Procedure
This section outlines the steps for calibrating the transmitter’s attached
sensors.
Note:
Perform the zero calibration before the span calibration.
NOTE
!
1. If using a compressed gas cylinder, push the calibration gas
flow housing onto the bottom of the sensor and apply the gas.
2.
Access the Gas Calibration Menu. This menu is for both zero
and span calibrations.
Summary of Contents for XNX-AM Series
Page 2: ...This page is intentionally left blank ...
Page 7: ...XNX Universal Transmitter Safety and Information 5 Safety and Information ...
Page 12: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 10 Introduction ...
Page 27: ...XNX Universal Transmitter Introduction 25 Menu Navigation ...
Page 102: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 100 Calibration ...
Page 118: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 116 Maintenance ...
Page 122: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 120 Warnings and Faults ...
Page 139: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 137 Specifications ...
Page 174: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 172 Control Drawings ...
Page 177: ...XNX Universal Transmitter Control Drawings 175 ...
Page 179: ...XNX Universal Transmitter Control Drawings 177 Remote Sensor Mount ...
Page 180: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 178 HART Protocol ...
Page 193: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 191 Modbus Protocol ...
Page 200: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 198 Warranty ...
Page 202: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 200 Index ...