background image

 

21 

 

oxygen.  During the bump test the oxygen 
sensor should read /-0.5% 
(absolute) of the level given on the 
calibration cylinder. 

4.2 

Fresh Air/Zero Calibration 

  Fresh air/zero 

calibrations may only be performed in an 
atmosphere that is known to contain 20.9% 
oxygen, 0% LEL and 0 PPM toxic gas. 

To initiate the fresh air/zero calibration: 
1.  Press the MODE button three times within 

two seconds to begin the fresh air/zero 
calibration sequence.  The MultiPro will 
briefly display AUTO CAL and then begin 
a 5-second countdown. 

 

 

2.  Press the MODE button before the end of 

the 5-second countdown to begin the 
fresh air/zero calibration.  The fresh 
air/zero calibration has been initiated 
when the MultiPro alternates between the 
following two screens: 

 

 

 

2.  The fresh air/zero calibration is complete 

when the instrument begins another 5-
second countdown for the span 
calibration.  If span calibration is not 
required, allow the countdown to reach 0 
without pressing the MODE button. 

4.2.1  Fresh air/zero calibration failure 

In the event of a fresh 
air/zero calibration failure, 
the alarms will be 
activated and the 
instrument will display the 
following screen.  Note 
that the sensor(s) that fail 
the zero calibration are shown (in this case, 
CO) 
After 3 seconds, the MultiPro will return to the 
current gas readings screen and the visual 
and audible alarms will cease.   

When fresh air/zero calibration is due, the 
MultiPro’s display will show the warning 
symbol while intermittently displaying the 0-
CAL icon in the gas readings screen until a 
successful fresh air calibration is performed.  
If a specific sensor has failed the fresh 
air/zero calibration, it will be displayed with the 
0-Cal icon. 
If a successful fresh air / zero calibration is not 
performed prior to instrument shut down, the 
MultiPro will show that Fresh Air Calibration is 
due during instrument start up. 

Possible causes and solutions 

1.  The atmosphere in which the instrument is 

located is contaminated (or was 
contaminated at the time the instrument 
was last fresh air calibrated). 

2.  A new sensor has just been installed. 
3.  Instrument has been dropped or banged 

since last turned on. 

4.  There has been a significant change in 

temperature since the instrument was last 
used. 

Recommended action: 

Take the instrument to fresh air and allow 
readings to stabilize.  Perform the fresh 
air/zero adjustment again.  If the manual fresh 
air/zero procedure fails to correct the problem, 
perform the manual fresh air / zero calibration 
procedure as described in section 4.2.2 
below. 

4.2.2  Forced fresh air / zero calibration 

The MultiPro includes safeguards to prevent 
fresh air calibration in contaminated 
environments.  If the standard fresh air / zero 
calibration fails a second time, the instrument 
may be “forced” to accept the fresh air 
calibration by performing the manual fresh air 
/ zero calibration. 

  Fresh air/zero 

calibrations may only be performed in an 
atmosphere that is known to contain 20.9% 
oxygen, 0.0% LEL and 0 PPM toxic gas. 

1.  Initiate the standard 

fresh air / zero 
calibration sequence 
by pressing the 
MODE button three 
times in rapid 
succession.  The 5-
second countdown will begin. 

2.  Press the MODE button before the end of 

the 5-second countdown and

 continue to 

hold the MODE button

.  As in the 

Summary of Contents for MultiPro

Page 1: ......

Page 2: ... IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED MultiPro Reference Manual Part Number 13 278 Version 3 Copyright 2012 by Honeywell Analytics Inc Lincolnshire Illinois 60069 All rights reserved No page or part of this operation manual may be reproduced in any form without written permission of the copyright owner shown above Honeywell Analytics res...

Page 3: ...n using the MultiPro pump 11 1 8 Data storage 11 1 8 1 Black box data recorder 11 1 8 2 Event logger 11 1 9 MultiPro design components 11 1 10 MultiPro standard accessories 12 1 10 1 Alkaline MultiPro detectors 12 1 10 2 Li Ion MultiPro detectors 12 1 11 MultiPro kits 12 1 11 1 MultiPro Confined Space Kits 12 1 11 2 MultiPro Value Packs 12 2 BASIC OPERATIONS 12 2 1Turning the MultiPro On 12 2 1 1 ...

Page 4: ...lines for the Li Ion versions 24 5 5 2 Charging guidelines for Li Ion battery 25 5 5 3Charging procedure for Li Ion battery 25 5 5 4 Charging with the pump attached 25 5 5 5 Battery troubleshooting 25 5 6 Sensor replacement 25 5 6 1 Sensor replacement Not Duo Tox 25 5 7 Sample probe assembly 26 5 7 1 Changing sample probe filters 27 5 7 2 Changing sample probe tubes wands 27 5 8 MultiPro Pump Main...

Page 5: ...Appendix D Toxic Sensor Cross Sensitivity 33 Appendix E Basic Parts List 33 MultiPro Remote Sampling Accessories 33 MultiPro Sensors 33 MultiPro Charging Datalogging Accessories 33 Appendix F Honeywell Analytics Standard Gas Detection Warranty 35 ...

Page 6: ...entially hazardous situation which if not avoided could result in death or serious injury indicates a potentially hazardous situation which if not avoided may result in moderate or minor injury CAUTION used without the safety alert symbol indicates a potentially hazardous situation which if not avoided may result in property damage Warnings and Cautions 1 The MultiPro personal portable gas detecto...

Page 7: ...ere that is known to contain 20 9 oxygen 0 0 LEL and 0 PPM toxic gas 11 The accuracy of the MultiPro should be checked immediately following any known exposure to contaminants by testing with known concentration test gas before further use Failure to check accuracy can lead to inaccurate and potentially dangerous readings 12 The accuracy of the MultiPro should be checked immediately following any ...

Page 8: ...eference manual before operating or servicing the MultiPro 18 A rapid up scale reading followed by a declining or erratic reading may indicate a hazardous combustible gas concentration that exceeds the MultiPro s zero to 100 percent LEL detection range 19 The MultiPro is not designed for use in oxygen enriched atmospheres 20 Do not use the MultiPro pump for prolonged periods in an atmosphere conta...

Page 9: ... the end of the sample draw probe Use of the hand aspirated sample draw kits is covered in section 3 1 Use of the motorized sample draw pump is covered in section 3 2 A detailed description of the MultiPro probe assembly is given in section 5 7 1 2 Multi sensor capability The MultiPro can be configured to simultaneously monitor oxygen carbon monoxide hydrogen sulfide and combustible gases and vapo...

Page 10: ... to immediately leave the affected area and to return only after further testing determines that the area is once again safe for entry Failure to immediately leave the area may result in serious injury or death A rapid up scale reading followed by a declining or erratic reading may indicate a hazardous combustible gas concentration that exceeds the MultiPro s zero to 100 percent LEL detection rang...

Page 11: ...ously for proper voltage The MultiPro also monitors the connection of sensors that are currently installed The detection of any electronic faults causes the activation of the audible and visible alarms and causes the display of the appropriate explanatory message 1 5 1 Heartbeat At the center of the display is a heart symbol that will blink every few seconds to show that the instrument is function...

Page 12: ...ro will begin to write the new data over the oldest data The black box data recorder will store approximately 41 hours of data in one minute increments before it begins to write new data over the oldest data In this way the newest data is always conserved To extract the information from the black box data recorder the MultiPro must be returned to Honeywell Analytics Once the data is downloaded fro...

Page 13: ... ten feet of tubing The sample probe itself is also available separately Standard configurations of the MultiPro are delivered in a cardboard box with cardboard inserts 1 10 1 Alkaline MultiPro detectors If the MultiPro has been purchased as an alkaline instrument the standard accessories include a set of 3 disposable AA alkaline batteries 1 10 2 Li Ion MultiPro detectors If the MultiPro has been ...

Page 14: ...Pro instruments equipped with a pump see section 2 1 1 below For instructions on changing the time and date with the MODE button see section 6 below The instrument temperature will then be shown The warning alarm levels screen will then be shown followed by the danger STEL and TWA alarm levels screens For more information concerning STEL and TWA alarms see sections 2 4 3 and 2 4 4 After the alarm ...

Page 15: ...D once every 5 seconds as a reminder If the MultiPro is due for span calibration the calibration bottle icon will be shown Both 0 Cal and the calibration bottle icon will be shown if the instrument is due for both span and fresh air calibration If a sensor is not detected in one of the sensor channels during start up the reading in the designated sensor channel will appear blank If a complete sens...

Page 16: ...ence 2 4 3 STEL Alarms STEL Short Term Exposure Limit alarm values represent the average concentration of instrument readings for the target gas for the most recently completed 15 minutes of operation The default STEL alarm level for the MultiPro CO sensor is 100PPM The default STEL alarm value for the MultiPro H2S sensor is 15PPM 2 4 4 TWA Alarms TWA Time Weighted Average values are calculated by...

Page 17: ...ontaining 20 9 oxygen 0 LEL and 0 PPM toxic gases and then turned back on The display will show and OL alternately in place of the sensor reading for any channel that has gone into over range alarm A combustible sensor overrange alarm indicates a potentially explosive atmosphere Failure to leave the area immediately may result in serious injury or death In the event of an LEL overrange alarm the M...

Page 18: ...he bottom of the instrument The sensor ports must be kept free of obstruction Blocked sensor ports can lead to inaccurate and potentially dangerous readings In diffusion mode the atmosphere being measured reaches the sensors by diffusing through vents in the bottom of the instrument Normal air movements are enough to carry the sample to the sensors The sensors react quickly to changes in the conce...

Page 19: ...ds free remote monitoring Use of the motorized sample draw pump allows the MultiPro to continuously monitor remote locations The pump is powered by the MultiPro battery When the pump is attached to the instrument PUMP will be shown on the display in the current gas readings screen Note The maximum amount of tubing that can be used with the motorized sample draw pump is 50 feet To ensure accurate r...

Page 20: ...ump is designed to detect pressure changes while the sample draw probe is being held in a vertical position If the probe is held horizontally or at a low angle while inserted into a fluid a pressure drop sufficient to cause the pump to shut down may not be generated and water could be drawn into the pump assembly causing damage to the pump sensors and internal components of the MultiPro CAUTION In...

Page 21: ...or further details on sensor stabilization requirements 2 Make sure the instrument is located in fresh air Figure 4 1 Bump Test Span Calibration set up 3 Verify that the current gas readings match the concentrations present in fresh air The oxygen O2 sensor should read 20 9 0 2 vol The readings for the LEL sensor should be 0 LEL and toxic sensors should read 0 or 0 0 parts per million PPM in fresh...

Page 22: ...en until a successful fresh air calibration is performed If a specific sensor has failed the fresh air zero calibration it will be displayed with the 0 Cal icon If a successful fresh air zero calibration is not performed prior to instrument shut down the MultiPro will show that Fresh Air Calibration is due during instrument start up Possible causes and solutions 1 The atmosphere in which the instr...

Page 23: ...lytics calibration gas cylinders typically contain approximately 18 0 oxygen In order to pass the span calibration the instrument must register an oxygen reading below 19 5 during span calibration See section 4 3 2 below if the oxygen sensor does not detect the drop in oxygen level and fails the span calibration The calibration is fully automatic from this point on Upon successful calibration of a...

Page 24: ...anner as the toxic and LEL sensors If the O2 sensor fails to register a reading below 19 5 during the span calibration the display will show O2 Too Low followed by the O2 CAL Error screen immediately after the failed calibration attempt Press MODE to acknowledge the warning and turn the instrument off If the oxygen sensor fails to register the drop in oxygen during the span calibration while being...

Page 25: ...dous area may impair intrinsic safety Use only Duracell MN1500 or Ultra MX1500 Eveready Energizer E91 LR6 Eveready EN91 batteries Substitution of batteries may impair intrinsic safety To replace the alkaline batteries 1 Remove the battery pack from the MultiPro as discussed in above in section 5 3 2 Remove the three alkaline batteries and replace them Be sure to align the positive and negative end...

Page 26: ...e charger and with the guide pins facing down towards the charger as shown in the illustration at right Once the spacer has been inserted the MultiPro with pump may be charged according to the instruction given in section 5 5 3 above 5 5 5 Battery troubleshooting If the green Charge LED on the charger fails to light when the MultiPro with Li Ion battery pack is placed in the charger remove the ins...

Page 27: ...d 10 assume that the sensor stabilization period has passed 9 If an oxygen sensor was replaced perform the Fresh Air Zero calibration as discussed in section 4 2 10 If a combustible LEL or a toxic sensor has been replaced perform both the Fresh Air Zero calibration and the Span calibration as discussed in sections 4 2 and 4 3 5 7 Sample probe assembly The MultiPro s sample draw probe is the standa...

Page 28: ...e tubes are exchanged or replaced before being returned to service 5 8 MultiPro Pump Maintenance MultiPro pumps are fairly maintenance free with the exception of the replacement of the pump filters on a regular basis 5 8 1 Replacing pump filters 1 Remove the two screws that hold the inlet port to the pump 2 Gently pull the dust filter holder free of the pump 3 Remove and replace the dust filter th...

Page 29: ...ubmenu is entered the MultiPro will automatically scroll through the options at 5 second intervals To change a setting for on to off or vice versa hold the MODE button down for 1 2 seconds Once the settings have been modified as needed click the MODE button three times in rapid succession to return to the Main Menu STEL and TWA Alarms are discussed in Appendix A The Security Beep is covered in sec...

Page 30: ...nu Note To enable the H2S TWA alarm setting of 1 6 PPM first enable the H2S decimal point as described above in section 6 2 The 1 6 PPM alarm value will then appear as an option in the TWA alarm setting for H2S 6 6 Set cal gas Once the set cal gas submenu has been reached the MultiPro will scroll through the various calibration gas values screens and the set default cal gas option Click the MODE b...

Page 31: ...val and Bump Test Reminder are used exclusively with the MultiPro IQ Express Dock 6 9 Setup accept Once the settings in any of the submenus have been changed as needed return to the Main Menu by clicking the MODE button three times in rapid succession Once in the Main Menu hold the MODE button down for three seconds to exit The MultiPro will provide a countdown to accept changes to the settings Cl...

Page 32: ...ects due to chronic exposure at lower levels that take the greatest toll of affected workers This is the situation with regards to smokers parking garage attendants or others chronically exposed to carbon monoxide in the workplace Exposure levels are too low to produce immediate symptoms but small repeated doses reduce the oxygen carrying capacity of the blood over time to dangerously low levels T...

Page 33: ...uidelines concerning calibration frequency Honeywell Analytics procedures closely follow these guidelines If your operating procedures do not permit daily checking of the sensors Honeywell Analytics recommends the following procedure to establish a safe and prudent accuracy check schedule for your Honeywell Analytics instruments 1 During a period of initial use of at least 10 days in the intended ...

Page 34: ...NH3 C2H4 C2H2 Carbon Monoxide CO 100 10 5 10 15 5 15 50 15 3 0 75 250 Hydrogen Sulfide H2S 0 5 100 20 2 20 20 60 0 2 0 0 0 n d n d Appendix E Basic Parts List MultiPro Remote Sampling Accessories Part No Description 54 49 102 Continuous sample pump for MultiPro Includes 10 of sample tubing and sample probe assembly UL CSA approved version 54 49 102 5 Continuous sample pump for MultiPro Includes 10...

Page 35: ...ropean 54 49 107 MultiPro Li Ion rechargeable battery pack UL CSA Approved version 54 49 107 5 MultiPro Li Ion rechargeable battery pack ATEX approved version European 54 26 0603 MultiPro BioTrak Software Kit Includes BioTrak software and manual 54 26 0605 Infrared communication device Serial IrDA Requires one available PC serial port ...

Page 36: ...FITNESS FOR A PARTICULAR PURPOSE HONEYWELL ANALYTICS WILL NOT BE LIABLE FOR LOSS OR DAMAGE OF ANY KIND CONNECTED TO THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY Instrument Accessory Warranty Periods Product s Warranty Period PHD6 2 years from date of purchase ToxiPro MultiPro 2 years from date of purchase Battery packs and chargers sampling pumps and other com...

Reviews: