7
urged to use only Honeywell Analytics calibration materials when calibrating the
MultiPro.
16.
Substitution of components may impair intrinsic safety.
17.
For safety reasons this equipment must be operated and
serviced by qualified personnel only. Read and understand this reference manual
before operating or servicing the MultiPro.
18.
A rapid up-scale reading followed by a declining or erratic
reading may indicate a hazardous combustible gas concentration that exceeds the
MultiPro’s zero to 100 percent LEL detection range.
19.
The MultiPro is not designed for use in oxygen enriched
atmospheres.
20.
Do not use the MultiPro pump for prolonged periods in an
atmosphere containing a concentration of solvent or fuel that may be greater than 50%
LEL.