58
APT4000PH
Configuration
Control of rinsing and calibration probes
Rinsing/calibration probe
Rinsing interval
Rinse duration
Contact response
Cleaning interval
Rinsing and cali-
bration probes
Pb.
Code Display
Menu group
Select menu item
Calibration interval
rinse
(EASYCLN)
EASYCLN:
see opposite
page
Code Display
Action (rinsing probe)
Selection
Pb.
Control of:
• Rinsing probe (rinse)
• Calibration probe (EasyClean)
Select with
key
Proceed with
ENTER
000.0 h
(
XXX.X h)
Rinsing interval
Select with
key, edit number
with key,
proceed
with
ENTER
Rinse duration
Select with
key, edit number
with key,
proceed
with
ENTER
0060 s
(0000 to
1999 SEC)
Contact type
Select with
key
Proceed with
ENTER
N/O
(
N/C)
59
EASYCLN
Code Display
Action (calibration probe)
Selection
Pb.
Calibration probe (EasyClean)
Select with
key
Proceed with
ENTER
Cleaning interval
(EasyClean only)
Select with
key, edit number
with key,
proceed
with
ENTER
000.0 h
(
XXX.X h)
000.0 h
(
XXX.X h)
Calibration interval
(EasyClean only)
Select with
key, edit number
with key,
proceed
with
ENTER
Off
(
On)
Lock cleaning (calibration)
interval*
On
: The APT4000PH only starts a
cleaning (calibration) interval if the
measured value lies within the toler-
ated range (Limit Lo/Limit Hi).
*“Lock cleaning (calibration) interval” function:
Limit Hi
Limit Lo
Locked
t
ppb
Locked
The APT4000PH only starts a
cleaning (calibration) interval if the
measured value lies within the
tolerated range (Limit Lo/Limit Hi).
(For limit setting, refer to Pages 53,
55)