Configuration
Limit function
Relay 1
Contact function
Contact response
Enter switching point
Enter hysteresis
Delay
Relay 2 menu group
Controller menu group
L1.
L2.
Ct.
Relay / Controller
rL.
Code Display
Menu group
Select menu item
End:
Press
CONF
, then
ENTER
53
LiMIT
(CtROL)
Code Display
Action
Choices
rL.
Use of relays:
• Limit function (LiMIT)
• Controller (CtROL)
Select with
key,
proceed with
ENTER
Note:
Selecting
CtROL
leads to
Controller menu group
Ct.
Lo
(Hi)
Limit 1 function , see Page 55.
Select with
key,
proceed with
ENTER
Limit 1 contact response
N/C: normally closed contact
N/O: normally open contact
Select with
key,
proceed with
ENTER
N/O
(N/C)
Limit 1 switching point
Select with
key, edit number with
key, proceed with
ENTER
00.00 pH
(
XX.XX pH
XXXX mV)
Limit 1 hysteresis
Select with
key, edit number with
key, proceed with
ENTER
00.50 pH
(
XX.XX pH
XXXX mV)
Limit 1 delay
The contact is activated with
delay (deactivated without delay)
Select with
key, edit number with
key, proceed with
ENTER
0010 SEC
(
0 to
9999 SEC)
L1.