![Honeywell APT4000PH User Manual Download Page 17](http://html.mh-extra.com/html/honeywell/apt4000ph/apt4000ph_user-manual_151787017.webp)
32
APT4000PH
Configuration
Output 1
Select type of electrode. Measurement procedure.
Select measured variable
Select electrode type
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA in the case of error
Hold mode
End:
Press
CONF
, then
ENTER
code
Output 1
o1.
Display
Menu group
Select menu item
33
Note:
Characters represented in gray are flashing and can be edited.
Code
Display
Action
Choices
o1.
Select configuration
(Press
CONF
.)
Enter mode code “1200”
(Select position with
key and
edit number with
key.
When the display reads “1200”,
press
ENTER
to confirm.)
The APT4000PH is in Hold mode
(Hold icon is on).
Select measured variable pH/ORP
Select with
key
Proceed with
ENTER
pH
/ORP
Only with pH selected:
Select electrode type:
• ISFET electrode
• Glass electrode
Select with
key
Proceed with
ENTER
FEt
(Glas)
After correct input a wel-
come text (ConF) is dis-
played for approx. 3 s