EN-
4
EN
Cleaning
Clean the Primair 700 using a mild cleaning solution [e.g. dish washing liquid soap or the EPI-US product (P/N 1779065)] and a lint-free cloth or sponge agent. Rinse the surfaces thoroughly with clean water
(do not use any solvents). During cleaning, ensure that no water is allowed to enter the air pump unit. If liquid gets inside the breathing hose, make sure it is thoroughly cleaned and dried before using it with the
RPD. Take extra care not to get cleaning solutions or water near the air-inlet, the battery connector contacts or the gasket. The apparatus must not be immersed in fluid.
Dry all components and the exterior of the breathing hose with a clean, lint-free cloth and/or leave to air dry in a clean environment free of dust and particulates. Make sure all components are completely dry before
putting the breathing device back into service.
MAINTENANCE __________________________________________________________________________________________________________________________________________________________________________________________
Maintenance, cleaning and replacement or repairs shall be done only by personnel familiar with these user instructions and power assisted filtering RPDs, using parts supplied by the original equipment manufacturer.
Prior to use:
Functional inspection of the system
After use:
Cleaning and disinfecting the complete power assisted filtering RPD.
Every 6 years:
Replace the breathing hose o-ring.
STORAGE_________________________________________________________________________________________________________________________________________________________________________________________________
Inspect all components prior to storage. Store the RPD in a clean and dry environment as its original packaging or in a sealed container so that it is protected from sunlight, dust or any contaminants. The recommended
storage temperature for Primair 700 Series is -20°C – +35°C and the relative humidity is less than 90%.
TECHNICAL SPECIFICATIONS
____________________________________________________________________________________________________________________________________________________________________________
Operating conditions:
Temperature:
-12°C – +48°C
Relative humidity:
20-90%
Storage conditions:
Temperature:
-20°C – +35°C
Relative humidity:
< 90%
Blower unit:
Ingress Protection:
IP65 with battery, breathing hose, filter and filter cover installed
Housing/material:
PC/ABS
Motor life:
>10 000 hours
Noise level:
<75dBA
Flow rate (tight-fitting config.):
145 lpm
Flow rate (loose-fitting config.):
180-195-210 lpm
Hose
connection:
bayonet
Weight:
805 g
Altitude:
0 – 3048 m above sea level
Explosive Atmosphere:
Do not use.
Connected Blower unit PA710CEU:
Radio frequency:
2.4 GHz
Battery:
Type:
Li-Ion rechargeable
Material:
Housing:
PC/ABS
Latch:
Nylon
Gasket:
Open-cell silicone foam
Weight:
410 g
Capacity:
10.9V, 6700mAh, 73wH
Charging time:
4.5 hours
Charging cycles:
> 500 cycles
Minimum battery duration (hours)*
Filter type
Loose-fitting interfaces
Tight-fitting interfaces
Default (low) flow
Medium flow
High flow
Nominal flow
P R SL
8
8
8
8
A2 P R SL
7-8
6.5 -7.5
6 -7.5
5.5 - 6.5
* Based on laboratory testing.
Battery Charger:
Display:
Real-time battery charge and health monitoring
Power Supply:
AC 100-240V, 1.5A 50-60Hz
Single-bay Charger: DC 19.0V, 2.37A
Belt and Comfort Pad / Back Pack Materials (Standard version):
Straps:
Woven Nylon
Back/Hip Support:
Thermoplastic Polyurethane, PC/ABS
Comfort Pad:
Closed Cell Polyethylene Foam, Nylon Fabric (TPE Coated)
Back pack:
Thermoplastic Polyurethane (TPU)
TROUBLE SHOOTING GUIDE
_____________________________________________________________________________________________________________________________________________________________________________
PROBLEM
POSSIBLE CAUSE
CORRECTION
Blower will not turn on
Battery is not sufficiently charged.
Replace battery with charged battery.
Low air-flow alarm is active
1) Filter has become saturated with contaminates.
2) Breathing hose is obstructed.
3) Cartridge air-inlets are obstructed.
1) Replace the spoiled filter.
2) Check the breathing hose for blockages or other obstructions.
3) Check to ensure the filter air-inlets are clear of obstructions.
Low air flow inside the respiratory interface
Breathing hose is disconnected or blocked.
Check breathing hose to ensure there are no blockages inside.
Confirm both ends of the breathing hose are properly connected.
Low air flow inside the respiratory interface, but no alarm
1) Breathing hose is partially blocked.
2) Respiratory interface is damaged.
1) Check breathing hose to ensure there are no blockages inside.
2) Replace the respiratory interface.
Excessive air flow inside the respiratory interface, but no alarm
Magnet is missing from the blower-side connector of the breathing
hose.
Replace the breathing hose.
Caution symbol on display is constantly illuminated (confirmed
no low battery or low flow alarm)
System fault.
Return the defective unit.
Battery does not hold a charge
1) Battery is nearing the end of its useful life or is damaged.
2) Battery charger is malfunctioning.
1) Replace the battery with a new one.
2) Replace the battery charger.
Blower unit is not able to pair with the SafetySuite (PA710CEU
only)
Blower unit does not emit the operational data wirelessly.
Visit
https://papr.honeywell.com/
for FAQ and other helpful
information, or contact our SafetySuite support team.