FIGURE 8: ATEC STX-2F
FIGURE 9: ATEC STX-FISCHER
Cleaning Instructions
Follow these instructions to clean the Stereotactic Adapter:
1. Upon completion of the biopsy procedure, immediately rinse the adapter with warm water.
2. Spray the adapter with a cleaner.
CAUTION:
The following products are NOT RECOMMENDED as they may damage the exposed components:
a. Bleach based cleaning agents.
b. Hydrogen Peroxide
c. Any cleaners/disinfectants with a pH of less than 4 or greater than 10
3. Allow the necessary time for the chosen cleaner to disinfect the components.
4. Rinse the adapter with warm water.
5. Wipe the adapter dry with a cloth or a paper towel.
6. Cycle hub retaining clamps, handpiece retaining clamp, cock and fire adapter.
7. Re-Clean the stereotactic adapter if any component does not perform properly.
Note:
There are many configurations of stereotactic systems available. The ATEC stereotactic adapter configurations
depicted in this manual may require a unique fitting to connect to your particular manufacturer/model. Please contact
Hologic Customer Support at 1-877-887-8767 if you believe this accessory, along with the appropriate instructions, was
omitted. If you are an international customer, please contact your distributor directly with questions, comments, and/or
technical service issues.