ATEC Handpiece and Eviva Biopsy Device Connections
1. Peel open the saline bag, remove the cap and place the bag on the saline bag hook on the right side of the console.
2. Peel open the sterile pack of the disposable handpiece and place the tray on the top of the console.
CAUTION:
Leave the protective sheath on handpiece tip.
3. Remove the terminal ends of the four lines from the handpiece tray (refer to figure 3).
a. Install the spike to the saline bag and insert the silicone section of the handpiece tubing into the pinch valve labeled
“Saline PV”.
b. Plug the
red
banded line into the receptacle on the console with the red ring.
1. Eviva Biopsy Devices designed for use with stereotactic (STX) guidance are compatible with the
red
receptacle
marked “US/STX” on the console.
2. ATEC handpieces designed for use with the stereotactic (STX) and ultrasound (U/S) guidance are compatible
with the
red
receptacle marked “US/STX” on the console.
3. ATEC handpieces designed for use with MRI guidance are compatible with the
red
receptacle marked “MRI”
on the console.
c. Plug the
black
banded line into the receptacle on the console with the
black
ring.
d. Attach the
blue
suction fitting to the horizontal side port marked “PATIENT” on the suction canister lid
FIGURE 3: VACUUM AND HANDPIECE ASSEMBLY AND CONNECTIONS
3d
3c
3b3
3a
3b1 & 2
2
1