Equipment Operating Environmental Conditions

 

An ambient temperature of +15 ºC to +40 ºC. (+59 ºF to +104 ºF)

 
 

A relative humidity range of 30% to 75%

 
Operating altitude: less than or equal to 3,657 meters (12,000 ft.) above sea level

Electromagnetic Emissions

1.  The ATEC Sapphire Console must be installed and put into service according to the guidance provided in these 

instructions to ensure its electromagnetic compatibility. Refer to the electromagnetic emissions and immunity tables in 

this document.

2. Portable and mobile RF communications equipment can affect the ATEC Sapphire Console. Refer to the 

electromagnetic immunity tables in the Specifications section for recommended separation distances.

3. WARNING: This equipment/system is intended for use by healthcare professionals only. This equipment/system may 

cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation 

measures, such as re-orienting or relocating the ATEC Sapphire Console or shielding the location.

4. The use of any cables or accessories other than those specified in these instructions may result in increased emissions 

or decreased immunity of the ATEC Sapphire Console.

5. The ATEC Sapphire Console should not be used adjacent to or stacked with other equipment. If adjacent or stacked 

use is necessary, the ATEC Sapphire Console should be observed to verify normal operation in the configuration in 

which it will be used.

GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS

The ATEC Sapphire Console is intended for use in the electromagnetic environment specified below. The customer or the 

user of the ATEC Sapphire Console should ensure that it is used in such an environment.

TABLE 6: ATEC Sapphire (120V) 

Emissions Test

Compliance

Electromagnetic environment - 

Guidance

RF emissions
CISPR 11

Group 1

The ATEC Sapphire uses RF energy only 

for its internal function. Therefore, its RF 

emissions are very low and are not likely to 

cause any interference in nearby electronic 

equipment.

RF emissions
CISPR 11

Class A

The ATEC Sapphire is suitable for use in all 

establishments other than domestic, and 

those directly connected to the public low-

voltage power supply network that supplies 

buildings used for domestic purposes.

Harmonic emissions
IEC 61000-3-2

Not Applicable

Voltage Fluctuations/flicker emissions

IEC 61000-3-3

Not Applicable

TABLE 7:  ELECTROMAGNETIC IMMUNITY

Immunity Test

EN/IEC 60601 Test Level

Compliance Level

Electromagnetic Environment - 

Guidance

Electrostatic discharge (ESD) 
EN/IEC 61000-4-2 

±6 kV Contact 
±8 kV Air 

±6 kV Contact 
±8 kV Air 

Floors should be wood, concrete or ceramic 

tile. If floors are covered with synthetic 

material, the relative humidity should be at 

least 30%.

30%

75%

+15 ºC

+40 ºC

Summary of Contents for ATEC SAPPHIRE

Page 1: ...Drive Marlborough MA 01752 USA www hologic com 877 887 8767 Hologic UK Ltd Unit 2 Link 10 Napier Way Crawley West Sussex RH10 9 RA UK Hologic Ltd Heron House Oaks Business Park Crewe Road Wythenshawe...

Page 2: ...Multicare and StereoLoc are trademarks and or registered trademarks of Hologic Inc and or its subsidiaries in the United states and or other countries All other trademarks registered trademarks and p...

Page 3: ...opsy and Excision System Sapphire Unit Check Out 14 ATEC Breast Biopsy and Excision System Sapphire Unit Operating Instructions 14 Performing a Biopsy Using MRI Guidance 14 Performing a Biopsy Using S...

Page 4: ...ay cause harmful interference to other devices in the vicinity However there is no guarantee that interference will not occur in a particular installation If the console does cause harmful interferenc...

Page 5: ...oduction This manual is written for medical personnel who will be responsible for operating the ATEC Sapphire unit It is extremely important that the operator read and thoroughly understand the conten...

Page 6: ...on the Stereotactic and Ultrasound Environment 1 ATEC handpieces designed for use with stereotactic STX and ultrasound U S guidance are compatible with the red receptacle marked US STX on the console...

Page 7: ...rture MR MR dedicated handpiece EVIVA 12 12 guage 12 12 cm long 20 20 mm aperture T Petite 13 13 cm long 10 10 cm long TABLE4 EVIVABIOPSYDEVICECOMPATIBILITY Biopsy Device Needle Guide Stereotactic Ada...

Page 8: ...14 12 OB ATEC RTFA Stereotactic STX Hologic MultiCare Platinum Hologic Stereloc II Hologic Affirm Siemens Fischer Mammotest Siemens Fischer Mammotest Lateral Approach Siemens Opdima GE Senographe DS a...

Page 9: ...cavity 6 ManualAspiration Button When placed in Manual Aspiration mode the Saline PV is closed and the inner cutting cannula is retracted In this mode the user can vacuum the biopsy cavity by depress...

Page 10: ...ection NOTE Refer to Eviva IFU for Eviva biopsy device operating instructions and warnings ATEC Power Cord Connection and Circuit Breaker CAUTION Grounding reliability can only be achieved when this e...

Page 11: ...ower Cord plug 4 InstructionsPlacardHook Attachment site for the instructions placard 5 ATECHospitalGradePowerCord Connect the ATEC Power Cord into the power cord receptacle on the side panel AC inlet...

Page 12: ...cap are secure and sealed to avoid vacuum leaks 3 Plug the vacuum line assembly on the console into the top port labeled VACUUM on the suction canister lid FIGURE2 ATECCANISTER NOTE The ATEC Canister...

Page 13: ...g the red banded line into the receptacle on the console with the red ring 1 Eviva Biopsy Devices designed for use with stereotactic STX guidance are compatible with the red receptacle marked US STX o...

Page 14: ...ystem Sapphire Unit Operating Instructions Using the System in MRI Stereotactic or Ultrasound Guidance 1 When the console is in Biopsy mode depressing the footswitch will activate the vacuum and opera...

Page 15: ...tissue select Lavage mode It is not necessary to depress the footswitch 5 To vacuum the cavity in Lavage mode disconnect the luer lock of Y valve to open the saline line to the vent to atmosphere resu...

Page 16: ...a fully 1 Re install protective sheath 2 Suction canister lid is not seated onto canister 2 Properly seat lid on canister 3 Suction canister is cracked 3 Replace with new suction canister 4 Large port...

Page 17: ...itting is not connected to the suction canister lid d Connect suction fitting to patient port marked patient on the lid e Handpiece suction fitting is not connected to the correct port on suction cani...

Page 18: ...zed or a period of ninety 90 days from the date of shipment whichever period is longer Disposable supplies shall be warranted to conform to published specifications for a period ending on the expirati...

Page 19: ...cy Action Inspect footswitch cord Weekly Verify that the 90 connectors are facing toward rear of the console Verify that the footswitch cord is wrapped in the counterclockwise CCW direction around the...

Page 20: ...tops Vacuum releases from thumb Vacuum Line Assembly Removal with the console powered off 1 Locate the white lock collar where the vacuum line assembly attaches to the console 2 Lift up on the vacuum...

Page 21: ...le Installation with the console powered off 1 Press the new power cord into the cord receptacle 2 With 1 Philips screwdriver tighten fastener located on bottom of retaining bracket Verification of In...

Page 22: ...ild detergent Wipe dry WARNING Do not immerse console in water Immersion in water will cause console damage and may cause electric shock or death ATEC Footswitch Clean the ATEC Footswitch thoroughly w...

Page 23: ...in Wide 63cm 37 in High 93cm 21 in Deep 53cm Footprint 525 in 3400 cm 525 in 3400 cm Weight 100 lbs 45kg 100 lbs 45kg Maximum Power 1320W 1200W Voltage 120V AC 240V AC Frequency 60 Hz 50 Hz Maximum C...

Page 24: ...for US STX handpiece 20 ft 6 10m for MRI handpiece Sterilization Gamma Irradiation Eviva Biopsy Device Handpiece Size 1 81 in 4 60cm High x 1 48 in 3 76cm Wide x 8 17 in 20 75cm long Weight 8 7 oz 24...

Page 25: ...ther equipment If adjacent or stacked use is necessary the ATEC Sapphire Console should be observed to verify normal operation in the configuration in which it will be used GUIDANCEANDMANUFACTURER SDE...

Page 26: ...power supply or battery Power Frequency 50 60Hz Magnetic Field EN IEC 61000 4 8 3A m 3A m Power frequency magnetic fields should be that of a typical commercial or hospital environment Conducted RF I...

Page 27: ...output power of the communications device Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 8...

Page 28: ...or on the order of a physician YYYY MM DD Expiration date is represented by the following YYYY represents the year MM represents the month DD represents the day MRI Conditional an item that has been...

Page 29: ...Note 1 B Body Note 2 F Floating applied part ATEC and Eviva biopsy devices Alternating Current Manually Reset Breaker Keep away from rain Dangerous Voltage WEEE Symbol indicates separate collection o...

Page 30: ...________________ Email __________________________________________________________________________ Clinical Education Specialist Name ___________________________________________________________________...

Page 31: ...Representative Information Name _________________________________________________________________________ Phone _________________________________________________________________________ Email _______...

Page 32: ...rther forward movement The cocking lever will return by spring action to its at rest position Note The stereotactic adapter can only be cocked if the handpiece retaining clamp is in the lock position...

Page 33: ...th warm water 5 Wipe the adapter dry with a cloth or a paper towel 6 Cycle hub retaining clamps handpiece retaining clamp cock and fire adapter 7 Re Clean the stereotactic adapter if any component doe...

Page 34: ...ereoloc are trademarks and or registered trademarks of Hologic Inc and or its subsidiaries in the United States and or other countries All other trademarks registered trademarks and product names used...

Page 35: ...35...

Page 36: ...36 Hologic Inc 250 Campus Drive Marlborough MA 01752 USA www hologic com 877 887 8767...

Reviews: