54 HIT-4 Communication Program
HP-329
HIT-4L
Pulse Width:
Sets the pulse width in mS. Valid entries 4mS to 300mS.
Pulse Scale:
Pulse scaling that represents the number of output pulses per
least significant digit of displayed total determined by the total
decimal selection. Valid entries 0.01, 0.1, 1, 10 and 100.
Inputs Page
The Inputs Configuration screen is used to configure the Temperature and
Pressure inputs and select the Reference Conditions.
Temperature Input Type
:
Drop down menu selection:
RTD: 100 Ohm, DIN385
4-20mA: 4 to 20mA temperature transmitter
Temperature Units
:
Drop down menu selection:
C: Degrees Celsius
F: Degrees Fahrenheit
Temperature Min:
Temperature value for 4mA input. Valid entries -450F to
999F (-267.8C to 537.2C).