32 Configuration
HP-329
HIT-4L
INPUT MENU
Menu Item
Description
Options
Min
Max
Value
Default
TYPE T
Temperature
Input Type
RTD
4-20mA
N/A
4-20mA
MIN T
4 mA
Temperature
Numeric
Entry
-450 F
999 F
0 F
MAX T
20 mA
Temperature
Numeric
Entry
-449 F
1000 F
100 F
DEF T
Default
Temperature
Numeric
Entry
-450 F
1000 F
68 F
REF T
Reference
Temperature
Numeric
Entry
-450 F
1000 F
68 F
MIN P
4 mA Pressure
Numeric
Entry
0 psia
49999 psia
0 psia
MAX P
20 mA Pressure
Numeric
Entry
1 psia
50000 psia
500 psia
DEF P
Default Pressure
Numeric
Entry
0 psia
50000 psia
500 psia
REF P
Reference
Pressure
Numeric
Entry
0 psia
50000 psia
14.696
psia
DEF D
Default Density
Numeric
Entry
0.001 lb/gal
100.000 lb/gal
1.0 lb/gal
REF D
Reference
Density
Numeric
Entry
0.0001 lb/gal
100 lb/gal
1.0 lb/gal
4.2 Default Configuration
HIT-4L is fully configured by the factory prior to shipment. When the
instrument is purchased with a Hoffer Flowmeter or when calibration
and configuration data are supplied, the instrument is configured as
specified. When calibration or configuration data is not available, the
instrument is shipped with default values. Refer to the above table for
a listing of the HIT-4L factory default configuration.
Summary of Contents for HIT-4L
Page 2: ......
Page 30: ...24 Installation HP 329 HIT 4L This page intentionally left blank...
Page 42: ...Operation 36 HP 329 HIT 4L This page intentionally left blank...
Page 52: ...HIT 4 Communication Program 46 HP 329 HIT 4L This page intentionally left blank...
Page 68: ...62 HIT 4 Communication Program HP 329 HIT 4L This page intentionally left blank...
Page 71: ...Appendix A Declaration of Conformity 65 HIT 4L HP 329 APPENDIX A DECLARATION OF CONFORMITY...
Page 72: ...66 Appendix A Declaration of Conformity HP 329 HIT 4L...