HIT-4 Communication Program 55
HIT-4G
HP-328
Temperature Max:
Temperature value for 20mA input. Valid entries -449F to
1000F (-267.2C to 537.8C).
Temperature Default:
Select default temperature condition. Valid entries -450F to
1000F (-267.8C to 537.8C).
Temperature Calibrate 4mA
:
Connect 4mA source to temperature input and press 4mA button
to calibrate Min.
Temperature Calibrate 20mA
:
Connect 20mA source to temperature input and press 20mA
button to calibrate Max.
Pressure Units
:
Drop down menu selection:
PSIA
BAR-A
PSIG
BAR-G
Pressure Min:
Pressure value for 4mA input. Valid entries 0 psia to 49999 psia
(0 bar to 3447.3 bar).
Pressure Max:
Pressure value for 20mA input. Valid entries 1 psia to 50000 psia
(0.069 bar to 3447.4 bar).
Pressure Default:
Select default pressure condition. Valid entries 0 psia to 50000
psia (0 bar to 3447.4 bar).
Pressure Calibrate 4mA
:
Connect 4mA source to pressure input and press 4mA button to
calibrate Min.
Pressure Calibrate 20mA
:
Connect 20mA source to pressure input and press 20mA button
to calibrate Max.
Summary of Contents for HIT-4G
Page 2: ......
Page 30: ...24 Installation HP 328 HIT 4G This page intentionally left blank...
Page 42: ...Operation 36 HP 328 HIT 4G This page intentionally left blank...
Page 52: ...46 Modbus Communications HP 328 HIT 4G This page intentionally left blank...
Page 68: ...62 HIT 4 Communication Program HP 328 HIT 4G This page intentionally left blank...
Page 71: ...Appendix A Declaration of Conformity 65 HIT 4G HP 328 APPENDIX A DECLARATION OF CONFORMITY...
Page 72: ...66 Appendix A Declaration of Conformity HP 328 HIT 4G...