44 Modbus Communications
HP-328
HIT-4G
Register
(Decimal)
Description
Data
Type
Access
Notes
8025
Calibrate Temp
Low
U16
R/W Write 5000 to calibrate
Read for A/D counts
8026
Calibrate Temp
High
U16
R/W Write 5000 to calibrate
Read for A/D counts
8027
Calibrate Press
Low
U16
R/W Write 5000 to calibrate
Read for A/D counts
8028
Calibrate Press
High
U16
R/W Write 5000 to calibrate
Read for A/D counts
8029
Temperature
Units
U16
R/W 0 == Fahrenheit;
1 == Celsius
8030
Pressure Units
U16
R/W 0 = PSIA, 1 = bar-a,
2 = PSIG, 3 = bar-g
8031
Density Units
U16
R/W 0 = lb/ft3; 1 = kg/m3
8038-39
†
Request Flow
Temperature
(16-bit)
FP32
RO
Flowing Temperature
in two consecutive
16-bit registers
8042-43
†
Request Flow
Pressure
(16-bit)
FP32
RO
Flowing Pressure in
two consecutive 16-
bit registers
8046-47
†
Request Flow
Compressibility
(16-bit)
FP32
RO
Flowing
Compressibility in
two consecutive 16-
bit registers
*
The Total and Grand Total values are available in 3 formats:
64-bit in a single register, 64-bit in four consecutive 16-bit
registers and 32-bit in two consecutive 16-bit registers. The
64-bit format is recommended for greater precision, especially
with values greater than 7-digits.
†
Temperature, Pressure and Compressibility are available in 16-
bit registers in firmware versions 1.020421 and later.
Summary of Contents for HIT-4G
Page 2: ......
Page 30: ...24 Installation HP 328 HIT 4G This page intentionally left blank...
Page 42: ...Operation 36 HP 328 HIT 4G This page intentionally left blank...
Page 52: ...46 Modbus Communications HP 328 HIT 4G This page intentionally left blank...
Page 68: ...62 HIT 4 Communication Program HP 328 HIT 4G This page intentionally left blank...
Page 71: ...Appendix A Declaration of Conformity 65 HIT 4G HP 328 APPENDIX A DECLARATION OF CONFORMITY...
Page 72: ...66 Appendix A Declaration of Conformity HP 328 HIT 4G...