Hitachi High-Tech Analytical Science
►
15
4 Description of the instrument
4.3
Technical data
Tab. 4-1
Technical data of the
Parameter
Value
Designation
Dimensions
width x height x depth
450 mm x 550 mm x 800 mm
Weights
tabletop unit
approx. 100 kg
max. sample weight
5 kg
Electrical system
supply voltage
100–240 V AC, 50–60 Hz
fusing
2 x 8 A slow-blowing, 5 x
20 mm
power consumption, max.
430 W
power consumption in idle state
(without PC)
50 W; source switched on
45 W; source switched off
protection class
IP20
power cable, IEC power
connector (for non-heating
appliances):
length
cross-section
2 m
3 x 1.5 mm
2
Optical system
multi-sensor optics
focal length
400 mm
number of grid lines
2400/mm
wavelength range
119–766 nm
Excitation source sparking
parameters
high-energy pre-spark (HEPS)
frequency
80–1000 Hz
voltage
up to 500 V
Argon
inlet pressure
3 bar
Ambient conditions
temperature
10–40 °C
air humidity
10–90 %, non-condensing
System requirements for
external PC
Operating system: The current Microsoft
®
Windows™ version at the
time of printing or the Microsoft
®
Windows™ version officially
designated by Hitachi High-Tech Analytical Science at the time of
ordering/delivery or its successor version.
Intel
®
Pentium
®
4, AMD Athlon™ 64 or better
8 GB RAM or more
100 GB available space
mouse
screen resolution: 1024 x 768 or better
1 x free USB 2.0 port
DVD drive
Options
wire adapter set
accessories set
consumables set
Summary of Contents for OE700 Series
Page 1: ...Operating manual 10022080 Ver 1 0GB 2019 08 14 OE700 series ...
Page 4: ......
Page 8: ...Table of figures VI Hitachi High Tech Analytical Science ...
Page 9: ......
Page 10: ...Table of figures VIII Hitachi High Tech Analytical Science ...
Page 14: ......
Page 20: ......
Page 26: ......
Page 36: ......
Page 46: ......
Page 48: ......
Page 62: ......
Page 66: ......
Page 68: ......
Page 70: ......
Page 72: ......
Page 74: ......
Page 76: ......