9 Maintenance
Adjusting the
electrode
6.
Completely press down the new electrode with the head (1, fig. 9-5) of the spacer
to adjust the correct distance to the sample. Simultaneously fasten the grub screw
to the electrode clamp.
Fig. 9-5
Installing the electrode
Spark-over!
If the spacer is not screwed in, this may cause a spark-over between the electrode
and the spark stand during sparking.
•
Never
spark without the spacer screwed into the spark stand!
7.
Screw the spacer back into the side of the spark stand finger-tightly.
1