9 Maintenance
►
40 Hitachi High-Tech Analytical Science
9.2
Maintenance table
In the following table,
the abbreviation
means that the maintenance activity must be
performed
OH .............................................after x operating hours
m ...............................................after x months
y.................................................after x years
.
Tab. 9-1
Maintenance intervals OE700 series
Assembly
Maintenance activity
Maintenance interval
Further information
m
y
OH
entire instrument replacement of all loaded parts and
analytical check of the instrument
1
to be performed by customer service
check of filters and replacement, if
required
3
cleaning the instrument
1
cleaning and filling the washing bottle
when dirty or filled to a low level
replacing the argon bottle
as necessary
spark stand
cleaning the spark stand
• after 1000 sparkings (to ensure
preferably precise analyses)
• upon notification
• at every change of matrix
replacing the spark stand plate
at every change of matrix
cleaning of electrode with brush
after every measurement
replacing the electrode
at every change of matrix
adjustment of the electrode
after every change of electrode
changing the ceramic insert
at every change of matrix
Summary of Contents for OE700 Series
Page 1: ...Operating manual 10022080 Ver 1 0GB 2019 08 14 OE700 series ...
Page 4: ......
Page 8: ...Table of figures VI Hitachi High Tech Analytical Science ...
Page 9: ......
Page 10: ...Table of figures VIII Hitachi High Tech Analytical Science ...
Page 14: ......
Page 20: ......
Page 26: ......
Page 36: ......
Page 46: ......
Page 48: ......
Page 62: ......
Page 66: ......
Page 68: ......
Page 70: ......
Page 72: ......
Page 74: ......
Page 76: ......