it sticks to one side, if it does
Hipac will need to visit site.
If the bar is sticking but there is no side to side movement in the bar then you will need to
check the channeling the door to make sure it’s not too tight on the alarm.
Hipac should not
need to visit site.
• Once you have seen the LED change to orange, hold the bar down for 10 seconds and the
LED will change to red and the alarm will sound.
If the LED changes to red but the alarm doesn’t sound, you need to speak to your staff attack
providers to make sure their system is wired correctly.
Hipac do not need to attend site
. If
the Led doesn’t cycle through the colours of green, amber, red as it should then the system
may be wired incorrectly. Please consult the Hipac O&M manual to make sure it’s wired as it
should be.
Please Note:
Once the alarm has been activated it will take approximately 30 seconds to
reset. If you keep the magnet to the LED, whilst it is still in alarm is will display red. If it has
gone back to clear then it has reset.