3.1 Pre-Operation Inspection
1
2
3
4
5
6
7
8
9
10
11
付録
索引
61
7.
Check to see if the insulation resistance
reading of the instrument is 20 M
.
8.
Stop insulation resistance measurement.
See 3.2.2 "Ending Measurement" (page 70).
9.
Short-circuit the tips of the clips of the
red and black test leads of the instru-
ment.
10.
Press the
key to see if the test volt-
age setting is [
5.00 kV
].
11.
Hold down the
key for more than
one second to start insulation resistance
measurement.
12.
Check to see if the insulation resistance
reading of the instrument is 0.00 M
.
If a problem exists, discontinue use of
the instrument.
Insulation resistance
Voltage