8.3 Error Display
1
2
3
4
5
6
7
8
9
10
11
付録
索引
159
Err02
When power is turned on for
the first time after the backup
battery is replaced, Err02
appears.
Reinstall the
battery.
If Err02 appears even after
the battery has been rein-
stalled, the life of the backup
battery has expired, the bat-
tery is faulty, or some other
cause exists.
This
requires
repair.
(Err03 to
Err05 may
be tempo-
rarily dis-
played
during dis-
charging
after mea-
surement,
but this does
not indicate
a malfunc-
tion.)
Err03
Voltage measurement error
has occurred.
Err04
Current measurement error
has occurred.
Err05
Temperature measurement
error has occurred.
Err06
The discharge circuit is faulty.
E11
Details:
The actual temperature for temperature
compensation exceeds the convertible
range or the reference temperature exceeds
the selectable range..
Action:
Perform temperature compensation within
the temperature ranges specified in the
tables in Appendix 4 "Temperature Compen-
Error
display
Details
Action