4.3 Temperature Compensation (TC)
1
2
3
4
5
6
7
8
9
10
11
付録
索引
95
6.
Press the
key to confirm the ref-
erence temperature.
[
TC
] lights up and the instrument enters
temperature compensation mode.
The LCD displays the resistance at the
reference temperature converted from
the measurement.
The bar graph shows the value before com-
pensation.
Bar graph
(Value before
compensation)
Digital value
(Value after
compensation)
Actual Measurement Temperature Reference Temperature
• If the resistance before compensation
exceeds the measurement range, it
cannot be converted and the LCD
displays [
- - -
].
• After the instrument is placed in
temperature compensation mode,
measurement or input of temperature and
measurement of insulation resistance
may be conducted.
• However, if the instrument is placed in
temperature compensation mode when
the temperature is not held (
is off), measure or enter temperature
before measuring resistance. You cannot
measure resistance first.