Appendix 4 Temperature Compensation Table
1
2
3
4
5
6
7
8
9
10
11
付録
索引
173
Table No.9
Source
IEEE Std 43-2000 Recommended Practice for Testing Insu-
lation Resistance of Rotating Machinery (U.S.A.)
Object under test
Rotating machinery
Selectable reference
temperature range
20 to 60°C, (Default 40°C)
Convertible range of
actual temperature
used for measurement
20 to 60°C
Compensation
formula
Converted to a resistance value at
the reference temperature using
the formula below and result dis-
played.
Rtref : Resistance after compen-
sation for reference tem-
perature of tref°C
Rt
: Resistance measured at
the temperature of t°C
tref
: Reference temperature [°C]
t
: Actual temperature used
for measurement [°C]
R
tref
0.5
tref t
–
10
Rt
=