90
Accuracy Table
10
Temperature
Thermocouple type
Range
Accuracy
*1
K
-40.0°C to 400.0°C
±0.5% rdg. ±2°C
-40.0°F to 752.0°F
*2
±0.5% rdg. ±3.6°F
• Overload protection: 1000 V DC/1000 V AC or 2×10
7
V • Hz (energized for 1 minute)
Current under overload: Steady state 15 mA or less, transient state 0.8 A or less
• The DT4910 Thermocouples (K) are used.
• The accuracy does not include the error of the DT4910 Thermocouples (K).
• Display update rate: 1 time/s (including disconnection check)
*1: In an environment where the temperature of the instrument is ±1°C and
stable, the accuracy is speci
fi
ed.
Standard contact temperature compensation stability time: 120 minutes (When
the instrument environmental temperature changes quickly from 50°C to 23°C
(122.0°F to 73.4°F).)
*2: The °F display is activated using a special instrument operation.
11
AC Current (clamp sensor)
Range
Accuracy (only the instrument)
*1
Conversion rate
40 Hz to 1 kHz
10.00 A
±0.9% rdg. ±3 dgt.
0.05 A/mV
20.00 A
±0.9% rdg. ±3 dgt.
0.10 A/mV
50.0 A
±0.9% rdg. ±3 dgt.
0.25 A/mV
100.0 A
±0.9% rdg. ±3 dgt.
0.5 A/mV
200.0 A
±0.9% rdg. ±3 dgt.
1.0 A/mV
500 A
±0.9% rdg. ±3 dgt.
2.5 A/mV
1000 A
±0.9% rdg. ±3 dgt.
5 A/mV
• Input impedance: 1 M
Ω
±20% or less, 1000 pF or less
• The 9010-50, 9018-50, or 9132-50 clamp-on probe is used.
• The accuracy does not include the error of the clamp-on probe.
• Crest factor: 3 or less