69
Checking the Relative Value/Performing Zero Adjustment
Performing zero adjustment
When performing zero adjustment, the condition of the test leads
varies depending on the measurement function.
Perform zero adjustment, referring to the table below.
Measurement function
V, A,
Ω
,
Condition of the test leads
Short circuit
Open
1
1
22
33
4
Red
Black
44
Example 1: Resistance measurement
1
Select the measurement
function.
2
Connect the test leads to
the measurement terminals.
3
Allow the test leads to
short circuit.
4
Press
for at least 1
second.
(After zero adjustment: 0.0
Ω
)
5
Measure the resistance.
3
3
Red
2
2
Black
1
1
44
Example 2: Capacitor measurement
1
Select the measurement
function.
2
Connect the test leads to
the measurement terminals.
3
Allow the test leads to
open.
4
Press
for at least
1 second.
(After zero adjustment: 0.000
μ
F)
5
Measure the capacitor.
www.
GlobalTestSupply
.com
Find Quality Products Online at: