83
Accuracy Table
5.3
Accuracy Table
Guaranteed accuracy period
1 year
Guaranteed accuracy period
after adjustment made by Hioki
1 year
Regulated power supply range
4.0 V ± 0.1 V or more (until the power shutdown)
Accuracy guarantee for
temperature and humidity
23°C ± 5°C (73.0°F ± 9.0°F), 80%RH or
less (no condensation)
Temperature characteristic
• “Measurement accuracy × 0.1/°C” is
added (excluding 23°C ± 5°C (73.0°F ±
9.0°F)).
• For resistance of DT4252 and DT4253
60.00 M
Ω
range, “Measurement
accuracy × 0.4/°C” is added (excluding
23°C ± 5°C (73.0°F ± 9.0°F)).
Other conditions
The accuracy of the two connected L4931
Extension Cable Sets (3 m) is guaranteed.
• rdg. (reading or displayed value): The value currently being measured and
displayed on the measuring instrument.
• dgt. (resolution): The smallest displayable unit, i.e., the input value that
causes the digital display to show a “1”.
1
AC voltage
Range
Accuracy
*1
Input impedance
40 to 500 Hz
Over 500 Hz to 1 kHz
6.000 V
±0.9% rdg. ±3 dgt.
±1.8% rdg. ±3 dgt.
11.2 M
Ω
±2.0%
100 pF or less
60.00 V ±0.9% rdg. ±3 dgt.
±1.8% rdg. ±3 dgt.
10.3 M
Ω
±2.0%
100 pF or less
600.0 V
±0.9% rdg. ±3 dgt.
±1.8% rdg. ±3 dgt.
10.2 M
Ω
±1.5%
100 pF or less
1000 V
±0.9% rdg. ±3 dgt.
±1.8% rdg. ±3 dgt.
10.2 M
Ω
±1.5%
100 pF or less