41
Abort Measurement
Syntax
Query
:ABORt
Description
Measurement is abort (forced termination).
:READ?
cannot be abort.
Example
:TRIG:SOUR EXT
:INIT:CONT ON
*TRG
:ABOR
Aborts a measurement.
:TRIG:SOUR EXT
:INIT:CONT ON
*TRG;*WAI
:ABOR
In this case, a measurement cannot be aborted because the instrument waits the
measurement to finish.
Note
An abort cannot be executed as the instrument waits until all prior commands finish if the
query is sent after a
*WAI
command.
An abort cannot be executed as the instrument waits until all prior commands finish if the
query is sent after a
*OPC
and
*OPC?
command.
Read Measurement Value and Set and Query Response Data
Syntax
Command
:MEASure:VALid
<MR0>
Query
:MEASure:VALid?
Response
<MR0>
<MR0> = 0 to 7(NR1)
MR: Measurement register (see below)
Description
Command
Query
Uses bits in <MR0> to specify the measurement parameters to be returned
in response to a query from a
FETCh?/:READ?
command.
1 is displayed at power-on.
Returns the measurement parameters specified using bits in <MR0> in
response to a query from a
FETCh?/:READ?
command.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
Unused
Unused Unused Unused Unused
Total
judgment
result
Judgment
result
Measured
value
Measurement register 0 (MR0)
Example
:MEAS:VAL 7
Sets to return the Measured value, Judgment result and Total judgment result.
:MEAS:VAL?
7
The Measured value, Judgment result and Total judgment result have been set to be
returned.