(6) Measurement range
Set and Query Measurement Range
Syntax
Command
:RANGe
<Measurement range>
Query
:RANGe?
Response
<Measurement range (Ω)>
Description
Command
<Measurement range (Ω)> = 0.0
~
120.0E-03(NR3)
•
0.0 ≤ <Measurement range (Ω)> ≤ 3.0E-03
Measurement range is set to 3mΩ.
•
3.0E-03 < <Measurement range (Ω)> ≤ 10.0E-03
Measurement range is set to 10mΩ.
•
10.0E-03 < <Measurement range (Ω)> ≤ 120.0E-03
Measurement range is set to 100mΩ.
Query
Returns the currently set measurement range as the following format.
<Measurement range (Ω)> = 3.0000E-3/10.0000E-3/100.000E-3
Example
:RANG 100.00E-03
Set the measurement range to 100mΩ.
:RANG?
100.000E-3
The measurement range has been set to 100mΩ.
(7) Sampling Speed
Set and Query Sampling Speed
Syntax
Command
:SAMPle:RATE
<V / Z>,
<Measurement speed>
Query
:SAMPle:RATE?
<V / Z>,
Response
<Measurement speed>
<V> = V measurement sampling
<Z> = Z measurement sampling
<Measurement speed> = FAST / MEDium / SLOW
Example
:SAMP:RATE V, MED
Set the V measurement sampling speed to MEDIUM.
:SAMP:RATE? V
MEDIUM
The V measurement sampling speed has been set to MEDIUM.