8.2 Event Registers
────────────────────────────────────────────────────
Event status register 0 (ESR0)
Bit 7
SE
Sampling End
Sampling ended after the end of the sampling count set by
the ":RTC:COUNT" command.
Bit 6
ST
Start Time
Start time is reached.
Bit 5
PE
Printer Error
A printer paper end, head up, or temperature out-of-range
status was issued.
Bit 4
FE
Floppy Error
A floppy disk write error, read error, or disk full status
occurred.
Bit 3
ST
Stop Time
Timer and real time processing finished.
Bit 2
IE
Interval End
Interval finished.
Bit 1
CE
Clamp Error
The clamp was disconnected or connected, or an operation
failure occurred.
Bit 0
UE
Harmonic sampling processing end
Sampling ended after the end of the sampling count set by the
"
:HARMonic:RTC
" command.
8.2
Event Registers
This section explains only event registers added for the 9605-01.
For the event status registers ESR2, ESR21 to ESR26, refer to Section
12.3.13, "Event Registers" (7) and (8) in the instruction manual of the 3194.
Event status register 0 (ESR0)
This register is used principally to monitor start and stop processing events.
The bit0 is added for the 9605-01.
For details, refer to ":HARMonic:RTC" command.
The following commands are used for reading the event status register 0, and
for setting the event status enable register 0 and for reading it.
Reading event status register 0
∗
ESR0?
Setting event status enable register 0
∗
ESE0
Reading event status enable register 0
∗
ESE0?