10 percent chlorine bleach/90 percent water solution
1. Wipe the device with a clean cloth slightly dampened with a 10 percent bleach and 90 percent
water solution.
2. Rinse with a clean cloth slightly dampened with clean water.
3. Allow the device surface to dry for a minimum of 10 minutes before using the device.
Clean the accessories
Wipe the NIBP hose and any reusable cuffs with a damp cloth moistened in a mild detergent
solution.
The same cleaning agents used to clean the device can be used on the stand and mounting
accessories.
WARNING
Electric shock hazard. If you clean the power transformer take the
following precautions:
•
Ensure the power transformer is unplugged from the outlet.
•
Ensure cleaning cloth is not dripping or over saturated
•
Apply minimal pressure on wipe when cleaning; avoid wiping gap or connection areas on the
power transformer and wall plug.
•
Ensure the power transformer, wall plug, and cord are dry before plugging into outlet.
NOTE
Clean the device on a routine basis, according to your facility's protocols
and standards or local regulations.
Factory defaults
Setting
Default value
Bluetooth
Enable
Unit of measure
mmHg
Pressure presets
160 mmHg
Data management
Auto overwrite
MAP
On
Language
English
Disassembly and repair reference
Screws
The following table lists torque specifications for all screws.
56 Appendices
Welch Allyn® Connex® ProBP™ 3400 digital blood pressure device
Summary of Contents for Welch Allyn Connex ProBP 3400
Page 8: ...4 Symbols Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 12: ...8 Safety Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 20: ...16 Overview Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 22: ...18 Controls and indicators Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 24: ...20 Connections Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 26: ...22 Power up sequence Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 42: ...38 Disassembly and repair Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 46: ...42 Reassembly Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 64: ......