2. Confirm that the isolation bumpers on the side panels are in the first and third holes from the
top, as shown in the diagram below.
Separate the chassis assembly from the circuit board
assembly
Before you begin, remove the following as described in this manual:
•
Battery
•
Rear housing
•
Front housing and side panels
1. Disconnect the pump power connector from J80 on the blood pressure board.
2. Disconnect the step valve manifold from the step valve on the blood pressure board.
Service Manual
Disassembly and repair 31