Troubleshooting
This section lists symptoms that you might observe, possible causes, and suggested actions that
might eliminate the problem. These tables can help you diagnose and fix a problem. They do not
replace basic troubleshooting skills. You must still trace the source of the problem to the board or
module level to decide the best course of action. Welch Allyn does not support component-level
repair to the board or module. For available replacement parts, see “Field-replaceable units.”
The Welch Allyn Service Tool enables you to access the error log for this device. The error log
contains diagnostic information in the form of text string errors. You can find descriptions of the
text string errors in the service tool Help files, along with potential corrective actions that do not
require device disassembly. If these corrective actions do not solve the problem, reference the
observed behavior in the tables below.
NOTE
Numeric error codes displayed alongside the text string errors are for
engineering evaluation purposes. These codes are not intended for diagnostic use.
WARNING
Do not perform troubleshooting on a device that is emitting smoke or
exhibits other signs of excessive overheating. Disconnect the device from AC
power and call Welch Allyn Technical Support immediately.
CAUTION
Replace parts, components, or accessories only with parts supplied or
approved by Welch Allyn. The use of any other parts can lead to inferior device
performance and will void the product warranty.
Power
Symptom
Possible cause
Suggested action
The device does not power on The battery is depleted
Plug in the device and allow
the battery to fully charge.
The battery is defective
Replace the battery with a
fully charged battery.
An internal connection is faulty
Verify that the battery contact
harness is not damaged, is
installed with the correct
polarity, and is securely
connected to J2 on the main
board.
23
Summary of Contents for Welch Allyn Connex ProBP 3400
Page 8: ...4 Symbols Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 12: ...8 Safety Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 20: ...16 Overview Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 22: ...18 Controls and indicators Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 24: ...20 Connections Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 26: ...22 Power up sequence Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 42: ...38 Disassembly and repair Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 46: ...42 Reassembly Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 64: ......