Symptom
Possible cause
Suggested action
Verify that the flex cable is
securely installed in J5 on the
main board.
The key panel is defective
Replace the front housing.
The main board is defective
Replace the main board.
The battery does not charge or
run time is low
The battery is defective
Verify all external connections
and allow the battery to
charge for 6 hours. If the
problem persists, replace the
battery.
An internal connection is faulty
Verify that the battery contact
harness is not damaged, is
installed with the correct
polarity, and is securely
connected to J2 on the main
board.
The main board is defective
Replace the main board.
Display
Symptom
Possible cause
Suggested action
The display is blank or cracked,
or the backlight is not on
The display is faulty
Replace the main board.
Communication
Symptom
Possible cause
Suggested action
A button does not respond
correctly
An internal connection is faulty
Verify that the flex cable is
securely installed in J5 on the
main board.
The key panel is defective
Replace the front housing.
The main board is defective
Replace the main board.
Cannot communicate via the
USB connection
The device is off
Verify that the device is
powered on.
The USB cable is faulty
Replace the USB cable.
Incorrect PC configuration
Test the connection by
connecting the device to a PC
running the service tool. Verify
that the service tool is
configured properly to
24 Troubleshooting
Welch Allyn® Connex® ProBP™ 3400 digital blood pressure device
Summary of Contents for Welch Allyn Connex ProBP 3400
Page 8: ...4 Symbols Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 12: ...8 Safety Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 20: ...16 Overview Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 22: ...18 Controls and indicators Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 24: ...20 Connections Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 26: ...22 Power up sequence Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 42: ...38 Disassembly and repair Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 46: ...42 Reassembly Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Page 64: ......