9. Align the inner chassis notches with the side panel guides, and align the dump valve manifold
with the pressure transducer on the blood pressure board.
10. Push the dump valve manifold onto the pressure transducer until fully seated.
11. Install the rear housing, battery, and battery compartment door by reversing the applicable
disassembly instructions.