135
136
SPECIFICATIONS
10
SPECIFICATIONS
10
PARAMETER
DESCRIPTION
Ventilator AC Battery Charger Specifications
Input AC voltage
100-240 VAC
Input AC frequency
50-60 Hz
Input AC current
0.3 A max.
Output DC voltage
8.4 VDC
Output DC current
1.3 A
Insulation class
Class II
Electrical safety approvals
UL 60601-1, EN 60950, EN 60601-1, EN 60335-2-29
Dimensions
3.55” x 1.77” x 1.26” (90 x 45 x 32 mm)
Weight
0.25 lb.. (115 g)
Ventilator Power Cord Specifications
Cord length
6 ft.
Compliance
IEC 320-E7
Compressor Power Supply Specifications
Input AC voltage
100-240 VAC
Input AC frequency
50-60 Hz
Input AC current
5 A max. at 100 VAC
2.5 A max. at 240 VAC
Output DC voltage
24 VDC
Output DC current
12.5 A
Insulation class
Class II
Ingress protection rating
IP22
Electrical safety approvals
IEC 60601-1, IEC 60601-1-11
Dimensions
7.8” x 4” x 2” (198 x 102 x 51 mm)
Weight
3 lb. (1.36 kg)
Compressor Power Cord Specifications
Cord length
6 ft.
Compliance
IEC 320-C13
Compressor Output
Pressure
58 PSI ±15% (at sea level and normal ambient conditions)
Flow
17
–
21 LPM (at sea level and normal ambient conditions)
Outlet Fitting
9/16" - 18 male threaded fitting
PARAMETER
DESCRIPTION
Storage and Transport Environment
Storage & Transport
Temperature
-20°C to +60°C (-4°F to +140°F)
Storage & Transport
Humidity
10% to 95% non-condensing
Storage & Transport
Altitude
1100 hPa to 624 hPa or -2290 ft. to +8000 ft. (-700 m to +2439 m)
Operating Environment
Operating Temperature
+5°C to +40°C (+41°F to +104°F)
Operating Humidity
10% to 95% non-condensing
Operating Altitude
1100 hPa to 624 hPa or -2290 ft to +8000 ft (-700 m to +2439 m)
NOTES:
• The performance of the compressor may degrade in high temperature, high humidity, or high altitude
environments. If degradation is seen, switch to an alternate means of ventilation.
• The ventilator settings might not be achieved
when sourced by the Life2000
®
Compressor
due to increases in altitude near or above 2500
feet. Consult the table to ensure that the
compressor can meet the ventilator settings.
The tidal volume delivered to the patient
includes the ventilator set entrainment
volume from patient int supplemental
oxygen volume (if used). For additional
Physical Features
Ventilation system weight
16 lb.
Ventilation system size
12.1” x 8.7” x 8.6”
Ventilator weight
1.1 lb.
Ventilator size
3.2” x 7.7” x 1.0”
Compressor weight
14.9 lb.
Compressor size
12.1” x 8.7” x 8.6”
Latex
The ventilation system does not contain natural rubber latex.
Expected Service Life
Expected Service Life
The Life2000® Ventilation System has a five year design life when operated in
accordance with the
Instructions For Use
provided.
Do not use the ventilation system past its expected service date.
The battery shall be replaced if performance degrades to an unacceptable level
during the one year warranty period.
Service Intervals
Service Intervals
1 year and 2.5 years from the date of shipment
Simulated Elevation
(in feet)
Observed Maximum
Compressor Output (in LPM)
0
17
2500
14
4000
12
8000
8