VENTILATION SETTINGS
6
VENTILATION SETTINGS
6
BREATH TYPES
There are two breath types that apply to the Volume Control ventilation provided by the ventilation system:
• Mandatory
• Assisted
MANDATORY BREATH
A mandatory breath (or machine breath) is completely controlled by the ventilation system. The system controls
both the beginning (triggering) and end (cycling) of the inspiratory phase.
ASSISTED BREATH
An assisted breath is controlled by both the patient and the ventilation system. Breaths are initiated by the
patient’s effort and volume delivery is controlled by the prescribed volume setting and inspiratory time.
VENTILATION MODES
The ventilation system delivers an inspired tidal volume to the patient according to the clinical settings: volume,
breath rate, trigger sensitivity, PEEP, and inspiratory time. Three different volume control modes are available:
• Control
• Assist/Control
• Assist
VENTILATION SETTINGS AND MODES
The
Ventilation Settings
screen provides options to set different ventilation modes. Consult the following table to
adjust parameters accordingly:
VENTILATION MODES*
SET TRIGGER SENSITIVITY TO:
SET BREATH RATE TO:
Control
OFF
≥ 1
Assist/Control
0 to 9
≥ 1
Assist
0 to 9
0
* SETTINGS BASED ON PRESCRIPTION
FACTORY DEFAULT PRESCRIPTION SETTINGS
The following table lists the Life2000® factory default
Prescription Settings
that may be edited. For a full list of
factory defaults, see the "Summary of Factory Default Settings" on page 77.
DESCRIPTION
DEFAULT VALUE
LOW
ACTIVITY
MEDIUM
ACTIVITY
HIGH
ACTIVITY
Ventilation Settings
Volume
150 ml
180 ml
200 ml
I-Time (Inspiratory Time)
0.75 sec
PEEP (Positive End Expiratory Pressure)
0 cmH
2
O
Sensitivity (Trigger Sensitivity)
4*
BR (Breath Rate)
12 BPM*
* Ventilation mode: Assist/Control ventilation mode
Alarm Limits
High BR (Breath Rate) alarm limit
50 BPM
Low BR (Breath Rate) alarm limit
5 BPM
High PIP (Peak Inspiratory Pressure) alarm limit
30 cmH
2
O
Low PIP (Peak Inspiratory Pressure) alarm limit
1 cmH
2
O
Breath Timeout
Breath Timeout Period
60 seconds
Breath Timeout Action
12 BPM
Source Gas
Source Gas
Air
NOTE:
Each
Prescription Setting
is independent of other
Prescription Settings
.
The following factory default cannot be edited:
DESCRIPTION
DEFAULT VALUE
High PEEP (Positive End Expiratory Pressure)
Pressure alarm limit
+7 cmH
2
O (above PEEP setting)