
117
118
MAINTENANCE
8
MAINTENANCE
8
TESTING SETUP
Undock the ventilator from the compressor. For more information, see "Undocking the Ventilator from the
Power on the ventilator. If necessary, record clinical settings before
changing settings.
Using the Clinician’s Menu, activate only the
Low Activity Prescription
Setting
.
Set the
Ventilation Settings
to:
Ventilation Settings
10:11a, Fri
Jun 12, 2015
ml
FLIP
CANCEL
OK
Volume
300
I-Time
PEEP
BR
sec
cmH
2
O
/min
0.75
0
12
0-9, OFF
Sensitivity
4
Touch
OK
and then
CONFIRM
.
Set the
Alarm Limits
to:
Set Alarm Limits
10:11a, Fri
Jun 12, 2015
/min
FLIP
CANCEL
OK
High BR
40
Low BR
High PIP
/min
cmH
2
O
0
40
cmH
2
O
Low PIP
1
Touch
OK
and then
CONFIRM
.
Set the
Breath Timeout
to:
Breath Timeout
20
Sec
12
BPM
10:11a, Fri
Jun 12, 2015
Touch
OK
and then
CONFIRM
.
Set the
Source Gas
to whatever source gas is being used for testing.
Connect the adjustable source gas supply to the
ventilator using an adjustable regulator set to 50
PSI. Turn on the source gas supply to 50 PSI
nominal.
Connect the interface. For more information, see
"Chapter 5: Connecting an Interface" on page 44.
NOTE:
The interface does not need to be worn during testing.
Place the ventilator on a flat surface where the interface connection will
not be jostled. Set the interface aside, with nasal pillows and tubing free
of anything that might impede air flow.
Prescription Settings
10:11a, Fri
Jun 12, 2015
3
2. TESTING THE LOW GAS PRESSURE ALARM (MEDIUM PRIORITY)
Begin by ventilating using the Low Activity button. Allow the ventilator to ventilate for a few breaths using the
current settings.
Disconnect the source gas supply hose from the ventilator connection by pulling back on the knurled ring
until the hose detaches.
The top of the touch screen should display the Low Gas Pressure alarm
(in yellow), and you should hear a sequence of three tones indicating a
medium-priority alarm or vibrate.
NOTE:
If a low- or medium-priority alarm occurs while the ventilator is in
vibration mode, and the alarm is not resolved within 60 seconds, an audible
alarm occurs.
After verifying the correct alarm notifications,
reattach the source gas supply hose to the
ventilator.
The Low Gas Pressure alarm will resolve itself
once the gas supply is correctly reattached.
3. TESTING THE HIGH DELIVERY PRESSURE ALARM (MEDIUM PRIORITY)
Begin by ventilating using the Low Activity button. Allow the ventilator to ventilate for a few breaths using the
current settings.
As soon as the breath indicator light shows a breath being delivered by the ventilator, quickly pinch the
interface tubing near the ventilator connection.
The top of the touch screen should display the High Delivery Pressure
(High Del. Pressure) alarm (in yellow), and you should hear a sequence of
three tones indicating a medium-priority alarm or vibrate.
NOTE:
If a low- or medium-priority alarm occurs while the ventilator is in
vibration mode, and the alarm is not resolved within 60 seconds, an audible
alarm occurs.
After verifying the correct alarm notifications, release the interface tubing.
The High Delivery Pressure alarm will resolve itself once the interface has
been released.
10:11a, Fri
Jun 12, 2015
18
Breath/min
O
2
LPM
PIP cmH
2
O
3.1
5.2
Low Gas Pressure
10:11a, Fri
Jun 12, 2015
14
Breath/min
O
2
LPM
PIP cmH
2
O
1.4
4.0
High Del. Pressure
Values displayed on the screen are for
illustrative purposes only.
Values displayed on the screen are for
illustrative purposes only.