42
Volara™ System, Model PVL1 User Manual (196286 REV C)
Volara™ Therapy
8.
Press the setting to adjust the pressure
setting.
9.
Use the + or – control to select the
pressure settings. Or,
use the
keypad
control to enter the required setting.
NOTE:
The maximum pressure that can be set depends on the pressure
ceiling limit defined in the
Clinical Access
settings. See page 49.
10. Press
Save
to save the changes.
Or, press
Cancel
.
•
If
Cough Pause
is enabled, press
Options
.
a.
Press the setting to adjust the
interval and duration of the cough
pause,
b.
Press the
keypad control and enter
the required setting, or use the
+
or
–
control to adjust the setting.
c.
Press
Save
to save the changes.
Or, press
Cancel
.
11. When you have completed the changes, press
Done
. The initial
therapy screen shows.
NOTE:
To prevent unauthorized persons from
adjusting the settings made, disable
Clinical Access
after you have made the
settings.
Press
Device
, then press
Control
. Slide
Clinical Access
to
OFF
.
GUI Map Rev 6 Page
46 TP1.13
196654__C_2019-Apr-29_Cleaned