49
Device Settings
S
ET THE
P
RESSURE
C
EILING
L
IMIT
The Pressure Management feature allows facility-authorized persons to
control the maximum pressure selection by setting a pressure ceiling
limit.The user is prevented from setting a therapy pressure that exceeds
the limit defined in the device settings.
WARNING:
Warning
—To avoid patient injury, this setting should only be configured
by facility-authorized persons.
NOTE:
To prevent unauthorized persons from
adjusting the settings made, disable
Clinical Access
after you have made the
settings.
1.
Press
Control
, then slide
Clinical
Access
to
OFF
.
2.
Press the
Home
menu control to
exit.
NOTE:
Available only if
Clinical Access
is enabled. See page 48.
1.
Press
Device
, and press
Controls
.
The default pressure ceiling shows on-
screen.