Volara™ System, Model PVL1 User Manual (196286 REV C)
33
Volara™ Therapy
Adjust Cough Pause Settings
NOTE:
Applicable only to preset therapy plans.
1.
Press
Automatic
.
•
If prompted, scan the patient’s ID
barcode with the paired barcode
reader. For pairing instructions, see
page 51.
1.
Before you start the therapy, swipe the
Options
tab left, and select the
Care
Plan
menu control.
2.
Select a preset therapy plan, and press
Options
.
3.
Slide
Cough Pause
to
ON
.
4.
To modify the interval and duration of the cough pause therapy,
press the setting.
a.
Press the
keypad control and enter
the required setting or,
use
+
or
–
controls to adjust the setting.
b.
Press
Save
to save the changes.
Or, press
Cancel
.
5.
When you have made all adjustments, press
Therapy
.
6.
Press
Start
. The therapy begins.