4.6 Duplex Receptacle
Chapter 4: Removal, Replacement, and Adjustment Procedures
CarePorter™ Cart Service Manual (man239)
Page 4 - 23
4
2. Using the screwdriver, tighten the terminal screws (G).
3. Push the duplex receptacle (D) into the opening in the device mount
weldment (E).
4. Using the phillips head screwdriver, install the two screws (C) that secure
the duplex receptacle (D) to the device mount weldment (E).
5. Cover the duplex receptacle faceplate (B).
6. Using the phillips head screwdriver, install the two screws (A) that secure
the duplex receptacle faceplate (B).
7. Test the duplex receptacle (D) for availability of power, and proper
grounding (refer to procedure 2.6).
Adjustment
No adjustment is required for the duplex receptacle (D).