7.4.7
Initializing the EEPROM
The following parameters/functions are reset to default values in the EEPROM.
Menu
Parameter / function Default
value
Runtime t/min,
t/sec
1:00
Speed RPM
1000
Relative centrifugal force
RCF 56
Centrifuging radius
RAD/mm
50
Brake stage
DEC fast
Acoustic signal after ending the centrifugation run.
End beep
on
Acoustic signal after a malfunction occurs.
Error beep
on
Volume of the acoustic signal.
Beep volume
max
Automatic unlocking of the lid after the centrifugation
run
Lid AutoOpen
on
Automatic shutdown of the backlighting
Power save
on
MACHINE MENU
Optical signal after ending the centrifugation run
End blinking
on
SERVICE MENU
Locking the keys
Key lock
Unlocked
Centrifuge model and type number
Version
10
Imbalance switch-off, limit value
Imbal limit
3500
Imbalance switch-off, PWM offset value
PWM offset
6500
SELECT MENU
Imbalance switch-off, threshold
Threshold
1400
After initializing the EEPROM, it is imperative that the following parameters/functions be set:
•
Centrifuge model and type (chapter 7.4.1)
•
Imbalance switch-off (see chapter 7.4.2).
With the rotor at a standstill, the EEPROM can be initialized as follows:
The procedure can be aborted at any time by pressing the
STOP
OPEN
key. In this case, the settings are not stored.
START
PULSE
+
STOP
OPEN
+
•
Keep both keys pressed and
switch on the mains switch at
the same time.
t/min:s
RPM
>RCF<
↓
SELECT
•
Press the key until the following is
displayed:
t/min:s
RPM
>RCF<
↓
Continued on next page