32/65
7.4.1
Setting the centrifuge model and type
The model and type number of the centrifuge must be set in the program.
The model and type number of the centrifuge are displayed briefly after the device is switched on.
With the rotor at a standstill, the model and type number of the centrifuge can be set as follows:
The procedure can be aborted at any time by pressing the
STOP
OPEN
key. In this case, the settings are not stored.
START
PULSE
+
STOP
OPEN
+
•
Keep both keys pressed and
switch on the mains switch at
the same time.
t/min:s
RPM
>RCF<
↓
SELECT
•
Press the key.
t/min:s
RPM
>RCF<
↓
START
PULSE
•
Press the key.
Centrifuge model.
Example:
t/min:s
RPM
>RCF<
↓
t
•
Set the centrifuge model with the keys.
Example:
t/min:s
RPM
>RCF<
↓
START
PULSE
•
Press the key.
Centrifuge type
Example:
t/min:s
RPM
>RCF<
↓
t
•
Set the centrifuge type with the keys.
Example:
t/min:s
RPM
>RCF<
↓
START
PULSE
•
Press the key to save the setting.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
↓
STOP
OPEN
•
Press the key once to exit the
"
Version" menu or twice to exit the
" SELECT MENU ".
t/min:s
RPM
>RCF<
Example:
t/min:s
RPM
>RCF<