Setting the imbalance
↓
If the imbalance setting is aborted
by pressing the
STOP
OPEN
key, the
following message appears.
Press any key to clear the message.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
↓
START
PULSE
•
Press the key.
The values of the imbalance system are
measured.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
If the system does not have
to carry out an offset
alignment:
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
If the system does have to
carry out an offset
alignment:
t/min:s
RPM
>RCF<
Example:
t/min:s
RPM
>RCF<
If values lie outside of the allowed
range, one of the following error
messages appears.
BaseVal too low :
Average value
too low.
BaseVal too high:
Average value
too high.
PWMadjust failed :
Offset
alignment failed.
Press any key to clear the error
message.
Possible cause:
•
Magnet holder not correctly
fastened to the motor.
•
Magnet missing or is defective.
•
The rubber/metal bearings are
worn.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
↓
Continued on next page