39
Name, location:
Address:
Place and postcode:
Station/room:
Name of the tester/date:
Functional testing:
No.
Description
Yes
No
Remark
Side rails:
21
Smooth running of the side rail in the tracks and safe locking into
place?
☐
☐
22
Sufficient mounting and firm seat of the side rail shafts/parts?
☐
☐
23
Load stress test of the side rails without deformation?
☐
☐
Lying surface:
24
Back part, leg part adjustment and special functions properly and
without any obstacles?
☐
☐
25
Safe grid mechanism of lower leg rest (if available) in every step,
even under stress?
☐
☐
26
Is the clamping effect of the 6 eccentric clamps (domiflex 2 /
domiflex 2 wash) sufficient? If not, the lock nut possibly could use a
bit more fastening!
☐
☐
Chassis (with scissors construction beds) / end panel (of actuator beds):
27
Height adjustment properly and without any obstacles?
☐
☐
28
Safe braking effect, blocking and free running of wheels?
☐
☐
Electric components:
29
Testing of hand control (keys and disabling function) all working
properly without any defects?
☐
☐
30
Testing of battery / block battery / emergency lowering: Function
properly and without any defects?
☐
☐
Electric measuring:
No.
Description
Yes
No
Measured
value
Insulation resistance -
(must be only measured on old models before manufacture year of 2002.)
31
Insulation resistance – measured value larger than 7MΩ?
☐
☐
Device leakage current –
(It must be measured with each TSC. Only it is a
domiflex 2 (item no.
271) or domiflex 2 wash
(item no. 272)
with a Limoss 24 V combinatorial circuit device (SMPS) drive system or a Dewert 24 V combinatorial circuit device (SMPS):
no leakage current measurement must be carried out during the service life of the health care bed when the visual and functional testing
(electric components) was passed.)
32
Device leakage current - measured value smaller than 0.5mA?
☐
☐
Evaluation
No.
Description
Yes
No
Remark
33
All values/inspection within the permissible range passed?
☐
☐
In the event the inspection result did not pass:
☐
Repair
☐
Sorting
Next inspection