15
(*) The comfort-sitting position only goes up. All adjusted position can be lowered by
pressing the button no.12
(**) When using the Trendelenburg function, the castors on one end panel have to be
unblocked. All castors have to be positioned parallel to be bed´s centreline.
(***) Not possible when using splitted (telescopic) side rails.
(1)
Back rest up
(2)
Back rest down
(3)
Ilumination below the lying surface
Please press button 1 + 2 simultaneously
(4)
Foot rest up
(5)
Foot rest down
(6)
Ilumination/reading lamp:
Please press button 4+5 simultaneously
(7)
Lying surface up
(8)
Lying surface down
(Intermediate stop at 37.5 cm exit height; safety stop at 27.5 cm)
(9)
Comfort-sitting position up
(*)
(10)
Foot down function (Anti-Trendelenburg)
(11)
Head down function (Trendelenburg)
(**)
(12)
Lying surface, back & foot rest down
(the bed goes down to the safety stop at 27.5 cm)
(13)
Ultra-low function:
Please press button 11 + 12 simultaneously
Attention contusion hazards of the feet
(Second safety stop at 15.5. cm for accessories and telescopic side
rail bars, final stop at 9.5 cm ultra low positon (***))
Reset: The reset process is activated by pressing sim-
ultaneously button 7 + 8. During this process, after 8
seconds, the bed goes slowly down into its lowest
position. You can hear a signal when the reset process
is fully completed.