11
The side rail is now in a low position.
If the side rails should be placed in the top position to aid in
fall prevention, reach with both hands in the centre of the
gripping groove in the upper ride rail and pull the side rail
upwards until you hear it click into place at both ends. The side
rail is now in a pulled-up position.
When using the telescopic (divided) side rails on this health care bed, it is necessary to
install four adjustable feet under the lying surface to prevent that the telescopic post of
the side rail sits on the floor. The screwed-on adjustable feet allow lowing of the bed
only to a minimum lower height of 15.5 cm.
Hazard note by Bock
Use only original Bock side rails, which are available as accessories for every health care bed.
Use only technically flawless and non-damaged side rails with the allowed gap dimensions.
Make sure that the side rails are engaged securely.
Before installation of the side rail and each new use, inspect all mechanical parts on the bed
frame and side rails, which secure the side rails for any possible damage.
The operation of side rails should be done with great care. Fingers can be quickly pinched be-
tween the longitudinal pieces.