
78
Guidance and manufacturer’s declaration – Electromagnetic immunity
The HemoCue systems are intended for use in the electromagnetic environment specified below. The customer or user of the
HemoCue systems should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output lines
2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±1 kV differential
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short interrup-
tions and
voltage variations on power
supply lines
IEC 61000-4-11
<5%U (>95% dip in U) for
a 0.5 cycle
40%U (60% dip in U) for
5 cycles
70%U (30% dip in U) for
25 cycles
<5%U (>95% dip in U) for
5 seconds
For explanation of U see
NOTE 1
<5%U (>95% dip
in U) for a 0.5 cycle
40%U (60% dip in
U) for 5 cycles
70%U (30% dip in
U) for 25 cycles
<5%U (>95% dip
in U) for 5 seconds
Mains power quality should be that of a typical
commercial or hospital environment.
If the user of the HemoCue systems image
intensifier requires continued operation during
power mains interruptions, it is recommended
that the HemoCue systems image intensifier be
powered from an uninterruptible power supply
or a battery.
Summary of Contents for WBC DIFF
Page 1: ...HemoCue WBC DIFF Operating Manual ...
Page 4: ...3 2 1 5a 7 4 6 5b 4 Components ...
Page 6: ...1 2 3 6 Start Up ...
Page 8: ...5 8 6 7 4 8 ...
Page 10: ...1 2 3 10 Data Entry Navigation ...
Page 18: ...1 2 3 4 5a 5b 18 Set Up Settings ...
Page 20: ...6 7 8 9 10 11 20 ...
Page 22: ...12 13 14 15 16 22 ...
Page 24: ...17 18 19 20 21 24 ...
Page 26: ...1 2a 2b 26 Set Up Printer ...
Page 28: ...1 28 Set Up Keyboard ...
Page 30: ...1 30 Set Up Barcode Reader ...
Page 32: ...1 32 Set Up Connectivity 2 ...
Page 34: ...1 3 2 34 Blood Collection Capillary Sample ...
Page 36: ...4 5 6 7 36 ...
Page 38: ...8 9 10 38 ...
Page 40: ...1 2a 2b 3 4 5 40 Blood Collection Venous Sample and Control Material ...
Page 42: ...1 2 4 3 5 42 Routine Use Patient Test ...
Page 44: ...6 7a 7b 8 44 ...
Page 46: ...10 9a 9b 9c 46 ...
Page 48: ...1 2 4 3 6 5 48 Routine Use QC Test ...
Page 50: ...50 7 8 9a 9b 10 ...
Page 52: ...12 11a 11b 11c 52 ...
Page 54: ...1 2 3 4 5 6 54 Review Delete Results ...
Page 56: ...7 8 9 56 ...
Page 58: ...1 2 3 4 5 58 Maintenance Daily Maintenance ...