72
Measuring Range
Displayed range for total WBC count: 0.3–30.0 x10
9
/L (300–
30000/mm
3
, 300–30000/µL) Results above the measuring range
will be displayed as HHH. Results below measuring range will be
displayed as LLL.
The differential cell count will be presented when the total white
blood cell count is between 1.0–30.0 x 10
9
/L.
Limit of Detection
Limit of Detection is defined as the lowest amount of analyte that
can be detected with probability.
Limit of Detection has been determined to be 0.3 x 10
9
/L. (CLSI
Document EP17-A) for WBC count.
Limitations of the Method/Procedure
a)
The measurement needs to be started no later than 1 minute
after filling the microcuvette.
b)
Do not remeasure a filled microcuvette.
c)
Mixing samples for an extended period may affect the result.
d)
Results above the measuring range will be displayed as
HHH. Results below the measuring range will be displayed
as LLL.
e)
Discard sample if clots.
f)
For interference studies see chapter
Known Interferences
.
Specific Performance Characteristics
Linearity
The white blood cell count for the WBC DIFF system has accord-
ing to the FDA Guidance Premarket Notification for Automated
Differential Cell Counters for Immature or Abnormal Blood
Cells, been demonstrated to be linear between 0.3 – 30.0 x 10
9
/L
with a correlation coefficient (r) to 0.999.
Capillary Sampling
When a capillary skin puncture is performed, several
defense systems in the body are activated very quickly.
These defense systems cause an increase in the number
of WBCs in the blood closest to the wound, leading to
greater differences in results from several samples taken
from the same finger stick.
Summary of Contents for WBC DIFF
Page 1: ...HemoCue WBC DIFF Operating Manual ...
Page 4: ...3 2 1 5a 7 4 6 5b 4 Components ...
Page 6: ...1 2 3 6 Start Up ...
Page 8: ...5 8 6 7 4 8 ...
Page 10: ...1 2 3 10 Data Entry Navigation ...
Page 18: ...1 2 3 4 5a 5b 18 Set Up Settings ...
Page 20: ...6 7 8 9 10 11 20 ...
Page 22: ...12 13 14 15 16 22 ...
Page 24: ...17 18 19 20 21 24 ...
Page 26: ...1 2a 2b 26 Set Up Printer ...
Page 28: ...1 28 Set Up Keyboard ...
Page 30: ...1 30 Set Up Barcode Reader ...
Page 32: ...1 32 Set Up Connectivity 2 ...
Page 34: ...1 3 2 34 Blood Collection Capillary Sample ...
Page 36: ...4 5 6 7 36 ...
Page 38: ...8 9 10 38 ...
Page 40: ...1 2a 2b 3 4 5 40 Blood Collection Venous Sample and Control Material ...
Page 42: ...1 2 4 3 5 42 Routine Use Patient Test ...
Page 44: ...6 7a 7b 8 44 ...
Page 46: ...10 9a 9b 9c 46 ...
Page 48: ...1 2 4 3 6 5 48 Routine Use QC Test ...
Page 50: ...50 7 8 9a 9b 10 ...
Page 52: ...12 11a 11b 11c 52 ...
Page 54: ...1 2 3 4 5 6 54 Review Delete Results ...
Page 56: ...7 8 9 56 ...
Page 58: ...1 2 3 4 5 58 Maintenance Daily Maintenance ...