59
Clean the cuvette holder after each day of use.
1. Turn off the analyzer. Place the cuvette moving arm in loading position.
2. Remove cuvette holder by lifting it straight up.
3. Clean the cuvette holder with alcohol (20-70%) or mild detergent.
NOTE: Do not autoclave.
If the optical parts are stained an error code is displayed. Clean according to
Maintenance/Optical Parts
section.
4. Wait 15 minutes before replacing the cuvette holder.
5. Place the cuvette moving arm in loading position before replacing the cuvette holder.
Clean the cover with alcohol (20-70%) or a mild detergent.
Only the WBC DIFF cuvette holder is to be used with the WBC DIFF analyzer.
The WBC DIFF cuvette holder is marked WBC DIFF.
Summary of Contents for WBC DIFF
Page 1: ...HemoCue WBC DIFF Operating Manual ...
Page 4: ...3 2 1 5a 7 4 6 5b 4 Components ...
Page 6: ...1 2 3 6 Start Up ...
Page 8: ...5 8 6 7 4 8 ...
Page 10: ...1 2 3 10 Data Entry Navigation ...
Page 18: ...1 2 3 4 5a 5b 18 Set Up Settings ...
Page 20: ...6 7 8 9 10 11 20 ...
Page 22: ...12 13 14 15 16 22 ...
Page 24: ...17 18 19 20 21 24 ...
Page 26: ...1 2a 2b 26 Set Up Printer ...
Page 28: ...1 28 Set Up Keyboard ...
Page 30: ...1 30 Set Up Barcode Reader ...
Page 32: ...1 32 Set Up Connectivity 2 ...
Page 34: ...1 3 2 34 Blood Collection Capillary Sample ...
Page 36: ...4 5 6 7 36 ...
Page 38: ...8 9 10 38 ...
Page 40: ...1 2a 2b 3 4 5 40 Blood Collection Venous Sample and Control Material ...
Page 42: ...1 2 4 3 5 42 Routine Use Patient Test ...
Page 44: ...6 7a 7b 8 44 ...
Page 46: ...10 9a 9b 9c 46 ...
Page 48: ...1 2 4 3 6 5 48 Routine Use QC Test ...
Page 50: ...50 7 8 9a 9b 10 ...
Page 52: ...12 11a 11b 11c 52 ...
Page 54: ...1 2 3 4 5 6 54 Review Delete Results ...
Page 56: ...7 8 9 56 ...
Page 58: ...1 2 3 4 5 58 Maintenance Daily Maintenance ...